• To avoid electric shock, unplug the device before cleaning it. DO NOT immerse the device in any fluids.
• Pins of connectors identified with the ESD warning symbol (
connectors unless ESD precautionary procedures are used. Precautionary procedures include methods to prevent build-up of
electrostatic charge (e.g., air conditioning, humidification, conductive floor coverings, non-synthetic clothing), discharging one's
body to the frame of the equipment or system or to earth or a large metal object, and bonding oneself by means of a wrist strap
to the equipment or system or to earth.
• Never operate the device if any of the parts are damaged, or if it is not working properly. Do not use the device if the water
chamber is leaking or damaged in any way. Have any damaged parts replaced before continuing use.
• Never touch the heater plate unless the unit is unplugged and the plate has cooled down.
• Allow the water in the chamber to cool to room temperature before removing the chamber from the humidifier.
• While the device is in operation, the soft flap on the flapper valve must move freely and close off the opening on top of the water
chamber. Replace the flapper valve if the flap is damaged or not intact.
• When installing the water chamber onto the system, do not allow any water to spill into the device.
• Empty and clean the water chamber daily to prevent mold and bacteria growth.
• When installing the water chamber onto the system, make sure the soft valve is first attached to the water chamber.
• If the device is used by multiple persons (e.g., rental devices), a low-resistance, main flow bacteria filter should be installed in-line
between the device and the circuit tubing to prevent contamination.
• The exhalation device (e.g., Whisper Swivel II) or exhalation port (on masks with an integrated exhalation port) is designed to
exhaust CO
from the patient circuit. Do not block or seal the ports on the exhalation device.
2
• If you are using a full face mask (a mask covering both your mouth and your nose), the mask must be equipped with a safety
(entrainment) valve. You must ensure that the entrainment valve is functioning properly.
• Route the wires to avoid tripping.
• This device is activated when the power cord is connected. Pressing the START/STOP button
(if activated) on or off.

Cautions

A Caution indicates the possibility of damage to the device.
• The device may only be operated at temperatures between 41° F and 95° F.
• If this device has been exposed to either very hot or very cold temperatures, allow it to adjust to room temperature (operating
temperature) before starting therapy. Do not operate the device outside of the operating temperature range shown in the
Specifications section.
• Do not immerse the device or allow any liquid to enter the enclosure, the inlet filter, or any opening.
• Do not place the device in or on any container that can collect or hold water.
• Make sure that the drain holes on the bottom of the device are not blocked.
• Condensation may damage the device. Always allow the device to reach room temperature before use.
• A properly installed, undamaged reusable foam inlet filter is required for proper operation.
• Tobacco smoke may cause tar build-up within the device, which may result in the device malfunctioning.
• If fluids are spilled onto the heater plate, unplug the power cord from the AC wall outlet and allow the device to drain and dry
before using.
• Take precautions to protect furniture from water damage.
• Use only distilled water in the chamber. The use of additives or chemicals in the chamber is prohibited.
• Do not fill water chamber above the fill line indicated on the side of the water chamber. Damage to the therapy device may
occur.
• Make sure the protective cover is securely fastened on top of the water chamber before operating the device.
• Do not place the device directly onto carpet, fabric, or other flammable materials.
• Do not allow the water chamber to sit for any length of time after it has been filled with water. Immediately install the chamber
in the device. Allowing water to sit in the chamber (when the chamber is not installed in the device) may cause the chamber to
separate from the bottom plate and may cause water leakage.
• Avoid moving the device when the water chamber has water in it.
• Operating the device with a dirty filter may keep the system from working properly and may damage the device.
• Dirty inlet filters may cause high operating temperatures that may affect device performance. Regularly examine the inlet filters as
needed for integrity and cleanliness.
• Never install a wet filter into the device. You must ensure sufficient drying time for the cleaned filter.
U
M
ser
anUal
)should not be touched. Connections should not be made to these
turns the airflow and humidifier
3