4 Operating the Power System

Programming the Output

Select a Channel View
Select an Output Channel
Set the Output Voltage
62
Front Panel:
Press the Meter key to toggle
between single-channel and
multiple-channel view.
Front Panel:
Press the Channel key to select an
output channel.
Front Panel:
Press the Voltage key.
Enter a value and press Select.
For models with multiple ranges, you can select a lower range if you
need better output resolution.
Front Panel:
Press the Voltage key.
Select a lower range and press
Select.
For Agilent Models N678xA SMU operating in current priority mode,
you can specify a Voltage limit., which limits the output voltage at the
specified value. In current priority mode, the output current remains
at its programmed setting as long as the output voltage is within the
positive or negative limit. Check Tracking to let the negative voltage
limit track the positive voltage limit setting.
Front Panel:
Press the Voltage key.
Specify a + Voltage limit and or a
− Voltage limit and press Select.
Check the Tracking box if you want
the − limit to track the + limit.
SCPI Command:
To select single-channel view:
DISP:VIEW METER1
To set multiple-channel view:
DISP:VIEW METER4
SCPI Command:
Enter the selected channel(s) in the
command's parameter list.
OUTP;STAT? (@1,2)
SCPI Command:
To set output 1 to 5 V:
VOLT 5,(@1)
To set all outputs to 10 V:
VOLT 10,(@1:4)
SCPI Command:
To select the lower range, program a
value that falls within the range:
VOLT:RANG 5,(@1)
SCPI Command:
To select the voltage limit:
VOLT:LIM 5,(@1)
To select the negative voltage limit:
VOLT:LIM:NEG 3,(@1)
To enable voltage limit tracking:
VOLT:LIM:COUP ON,(@1)