4 Operating the Power System
Set the Slew Rate
Enable the Output
64
Front Panel:
Select Output\Mode.
Select either Voltage priority or
Current priority and press Select.
When switching between Voltage priority and Current priority, the output is
NOTE
turned off and the output settings revert to their Power-on or RST values. Refer
to Appendix E for information about voltage and current priority mode operation.
Front Panel:
Select Output\Advanced\Slew
Then select Voltage or Current
Enter the voltage or current slew
rate in the Slew Rate field.
Check Max slew rate to program
the fastest slew rate.
The voltage slew rate determines the rate at which the voltage changes
to a new programmed setting. When set to MAXimum, INFinity, or to a
very large value, the slew rate will be limited by the analog
performance of the output circuit. The slowest or minimum slew rate
is model-dependent and is a function of the full scale range. For other
ranges the minimum slew rate is proportional to the full scale range.
The current slew rate control is only available on Agilent Models
N678xA SMU. It determines the rate at which the current changes to a
new programmed setting. When set to MAXimum, INFinity, or to a very
large value, the slew rate will be limited by the analog performance of
the output circuit. The slowest or minimum slew rate is model-
dependent and is a function of the full scale range. For other ranges
the minimum slew rate is proportional to the full scale range.
Front Panel:
Press the On/Off key.
To enable/disable ALL outputs
using the On/Off key, select
System\Preferences\Keys.
Check On/Off affects all channels.
Because of internal circuit start-up procedures and any installed
relay options, output on may take between 35 and 50 milliseconds to
complete its function. Conversely, output off may take between 20
and 25 milliseconds to complete its function.
To mitigate these built-in delays, you can program the output to zero
volts rather than using the output on/off function.
SCPI Command:
To set output 1 to current priority:
FUNC CURR,(@1)
SCPI Command:
To set the voltage slew rate to 5 V/s
VOLT:SLEW 5,(@1)
To set the current slew rate to 1 A/s
CURR:SLEW 1,(@1)
To set the fastest slew rate:
VOLT:SLEW MAX,(@1)
SCPI Command:
To enable only output 1:
OUTP ON,(@1)
To enable outputs 1-4:
OUTP ON,(@1:4)