Chapter 4 Remote Interface Reference
Simplified Programming Overview

Programming Ranges and Output Identifiers

Output setting commands require a parameter for programming ranges and an output
name or an output number as the identifier of each output and most queries will return
a parameter. The programming range for a parameter varies according to the selected
output of the power supply. The following table lists the programming ranges, output
names, and output numbers for each output.
Refer to this table to identify parameters when programming the power supply.
Table 4-1. Agilent E3631A Programming Ranges and Output Identifiers
Voltage
Programming
Range
MAX value
MIN value
*RST value
(DEFault value)
Current
Programming
Range
MAX value
MIN value
*RST value
(DEFault value)
Output identifier
Output number
74
Output
+6V output
+25V output
0 to 6.18 V
0 to +25.75 V
6.18 V
25.75 V
0 V
0 V
0 V
0 V
0 to 5.15 A
0 to 1.03 A
5.15 A
1.03 A
0 A
0 A
5 A
1 A
P6V
P25V
1
2
-25V output
0 to -25.75 V
-25.75 V
0 V
0 V
0 to 1.03 A
1.03 A
0 A
1 A
N25V
3