Connecting the Outputs

8A CONNECTOR
-
-
G +
+ +S G -S
INSERT
WIRES
TWIST
LEADS
-
+
LOAD
Series N6700 User's Guide
SHOCK HAZARD Turn off AC power before making rear panel connections.
WARNING
All wires and straps must be properly connected with the terminal block
screws securely tightened.
Disconnect the connector plug to make your wire connections. The
8A connector plug accepts wires sizes from AWG 14 to AWG 30. The
12A connector plug accepts wires sizes from AWG 12 to AWG 30. The
20A connector plug accepts wires sizes from AWG 10 to AWG 24. The
50A connector plug accepts wires sizes from AWG 6 to AWG 20. Wire
sizes smaller than AWG 20 are not recommended. Connect the load
wires to the + and - terminals. Connect the sense wires to the +s and
-s terminals. Sense jumpers are provided for local sensing.
Securely fasten all wires by tightening the screw terminals. Insert the
connector plug into the back of the unit. Secure the 12 A and 50 A
connector by tightening the locking screws. A chassis ground binding
post is located next to the AC input connector for ground
connections.
Part number information for all connector plugs is provided under
"Items Supplied" at the beginning of this chapter.
On power modules with the 50A sense connector, the +LS and –LS terminals
CAUTION
are ONLY used for local sense connections as illustrated. Do not connect these
terminals in any other way.
LOCKING
SCREW
+S
SENSE JUMPERS
INSTALLED FOR
LOCAL SENSING
+
TIGHTEN
SCREWS
12A & 20 A
CONNECTOR
LOCKING
SCREW
+
-S
INSERT WIRES
TWIST LEADS
LOAD
Installation 2
50 A
CONNECTOR
+
INSTALLED
FOR LOCAL
+
LOAD
50A
SENSE
+S +LS -LS -S
SENSE
JUMPERS
SENSING
31