2 Installation

Rack Installation

CAUTION
Tools required: Phillips driver, T22 Torx driver, T10 Torx driver
28
Step 3. Use a T10 Torx driver and install the screws at each end of the power
module. Because the RFI strips apply upward pressure, continue
pushing down on the module until the screws are tight.
Step 4. Replace the blower cover when finished. Carefully fit the spring clips
under the lip of the power modules.
You cannot use support rails for rack mounting your instrument.
Support rails would block the airflow needed for cooling.
Use Rack Mount kit (Option 908) to rack mount your instrument.
The Rack Mount Kit is also available by ordering part number N6709A.
Agilent N6700 MPS mainframes can be mounted in a 19-inch EIA
rack cabinet. They are designed to fit in one rack-unit (1U) of space.
Do not block the air intake and exhaust at the sides of the unit, or the
exhaust at the rear of the unit.
Step 1. Install eight clip-nuts on the rack frame (2 in each corner) where
your instrument will be located.
5
2a
1
4
1
6
2b
2c
Series N6700 User's Guide
3
2d