Logic Analyzer Test Strategy

Test Interval

Test Record Description

Test Equipment

Instrument Warm-Up

16800 Series Portable Logic Analyzers Service Guide
Only specified parameters are tested. Specifications are
listed on
page
15. The test conditions defined in this
procedure ensure that the specified parameter is as good as
or better than specification. No attempt is made to
determine performance which is better than specification.
Not all channels of the logic analyzer will be tested; rather a
sample of channels is tested. The calibration laboratory may
choose to elaborate on these tests and test all channels at
their discretion.
To perform a complete test on a logic acquisition module,
start at the beginning of the chapter and follow each
procedure.
Test the performance of the module against specifications at
two- year intervals.
For recording the results of each procedure, see
"Performance Test Record"
Each procedure lists the recommended test equipment. You
can use equipment that satisfies the specifications given.
However, the procedures are based on using the
recommended model or part number.
Before testing the performance of the module, warm- up the
logic analyzer and the test equipment for 30 minutes.
Testing Performance
on page 68.
3
27