6
Replacing Assemblies

To remove and replace the tray assembly

146
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 120.
"To remove and replace the cover"
"To remove and replace the front panel assembly"
page 129.
2
Using a Torx T10 screwdriver, remove the 2 side screws
securing the tray assembly to the chassis.
Slide the tray assembly out far enough to disconnect the
3
hard drive and power supply cables. Use the release
tab(s) to help disconnect the power supply cables.
Slide the tray out of the chassis.
4
T-10
screws (2)
Chassis
Hard
drive
cables
5
Reverse this procedure to replace the tray assembly.
16800 Series Portable Logic Analyzers Service Guide
on
on page 122.
on
Power supply
cables