6
Replacing Assemblies
To remove and replace a measurement card
138
Your product may consist of 1 or 2 measurement cards.
1
Perform previous procedures:
•
"To prepare the instrument for disassembly"
page 120.
Use a Torx T10 screwdriver to remove the 4 screws
2
securing the shroud to the frame.
3
Slide the shroud over the cables.
4
Use a gentle back- and- forth motion as you pull to
disconnect the card(s) from the connector on the module
interface board and then pull the card(s) out of the
frame.
5
Replace the measurement card(s).
16800 Series Portable Logic Analyzers Service Guide
on
Shroud
T-10
screws
(4)