Resistance Measurement Test - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

789
Calibration Manual

Resistance Measurement Test

1. Put the calibrator in Standby (
2. Put the UUT rotary switch in the
3. Connect the
4. Apply the calibrator resistance values in Table 8 in the 789 400 Ω to 40 kΩ range.
5. Change the connections to the UUT. Using the Fluke 5440A-7002 low thermal leads,
6. Apply the rest of the calibrator resistance values in Table 8 (400 kΩ range and
18
and
OUTPUT
solid and dotted lines in Figure 7.
Compare the readings on the UUT to the acceptable readings shown.
connect the calibrator to the UUT as shown by the solid lines in Figure 7.
above). Compare the readings on the UUT to the acceptable readings shown.
UUT
Figure 7. Resistance Measurement Test Connections
) mode.
STBY
V
position.
leads of the calibrator to the UUT as shown by the
SENSE
5500A
adm004F.EPS