Perform a Dual Flow Rate Test [CnD,a,b,c,d,e,f]

Fetch the Current Flow Measured on a Channel [FLOWn]

Fetch the Current Volume Accumulated on a Channel [VOLn]

B-4
This command starts a Dual Flow rate test on the specified channel. The IDA-4 Plus status
all channels screen will prompt the operator to prime the channel and start the test. Once
primed, the operator can press the start key on the front panel or send the [CnD] command
again.
Parameters:
a = control number of device under test
b = technician name
c = set primary flow rate
d = set primary volume
e = set secondary flow rate
f = set secondary volume
Returned to host:
[Flow, nnnn.nn,hh:mm:ss:mmm] <CR><LF>
Where:
nnn.nn is the flow rate in ml/hr
hh:mm:ss:mmm is the time since the start of the test, formatted as:
hh
Hours
ss
Seconds
mmm
Milli-seconds
Returned to host:
[Vol, nnnn.nn,hh:mm:ss:mmm] <CR><LF>
Where:
nnn.nn is the vol in ml
hh:mm:ss:mmm is the time since the start of the test, formatted as:
hh
Hours
mm
Minutes
mmm
Milli-seconds
n = channel number
mmMinutes
ssSeconds