D e v i c e
Control No
Operator
Flow Rate
Manufacturer
Device Type
Serial Num.
Location
ESC
FLOW
P r e s s E N T E R f o r n e x t f i e l d
P r e s s E S C f o r p r e v i o u s f i e l d
3-14
I n f o r m a t i o n
C h a n
1
12345
Your Name
10
Vol 20
OCCL
PCA
DUAL
Location
This is an optional field of sixteen characters in length.
It is used to store the normal location of the device
under test.
Pressing ENTER from this field advances to the SUB
MENU 2. ESC goes back to the Serial Number field

Sub Menu2

This menu allows you to select the type of test to
perform after you have entered all the information about
the device being tested.
ESC
This back tracks to the Location field. This gives the
operator a chance to correct any of the preceding
information that may have been entered incorrectly.
Flow
This advances to the FLOW TEST screen where the
operator will be guided through a flow rate test.
Occlusion
This advances to the OCCLUSION TEST screen
where the operator is prompted through the occlusion
test.
PCA
This advances to the PCA TEST INFORMATION
screen to request information required to perform a
PCA test.
Dual
This advances to the DUAL FLOW RATE
INFORMATION screen.