222
Reports are not printed or are incomplete
Probable cause 1:
Possible solution:
1
The printer is not supported by the QC-300.
Probable cause 2:
Possible solution:
1
Connect or replace the printer cable.
Probable cause 3:
Possible solution:
1
Correct the report setup in the Print setup screen.
Probable cause 4:
Solution:
1
Contact Metronics for a list of supported printers.
Reports are printed incorrectly
Probable cause 1:
Possible solution:
1
Correct the report setup in the Print setup screen.
Probable cause 2:
Possible solution:
1
Correct the control character configurations in the Print setup screen.
QC-300 Series User's Guide
The channel input device is not connected
or is malfunctioning
The USB cable is not firmly connected or
is damaged
The printed report setup is incorrect
The printer is not supported
The printed report setup is incorrect
The wrong control characters are specified