HEIDENHAIN GAGE-CHEK 2000 Series Operating Instructions Manual

HEIDENHAIN GAGE-CHEK 2000 Series Operating Instructions Manual

Evaluation unit
Table of Contents

Advertisement

Quick Links

GAGE-CHEK 2000
Operating Instructions
Evaluation Unit
English (en)
08/2020

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GAGE-CHEK 2000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for HEIDENHAIN GAGE-CHEK 2000 Series

  • Page 1 GAGE-CHEK 2000 Operating Instructions Evaluation Unit English (en) 08/2020...
  • Page 2 Setup..............................119 Measuring............................155 10 File management..........................169 11 Settings............................177 12 Servicing and maintenance......................211 13 What to do if ..........................221 14 Removal and disposal........................225 15 Specifications..........................227 16 Index...............................233 17 List of figures..........................235 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 3: Table Of Contents

    Intended use............................24 Improper use............................24 Personnel qualification........................25 Obligations of the operating company....................25 General safety precautions......................... 26 2.7 .1 Symbols on the product......................26 2.7 .2 Electrical safety precautions....................27 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 4 Overview............................... 42 General information..........................42 Device overview........................... 43 Connecting encoders........................... 45 Connecting touch probes........................46 Wiring switching inputs and outputs....................47 Connecting input devices........................48 Connecting a network peripheral.......................49 Connecting the line voltage........................49 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 5 6.11.1 Controls of the Inspector......................68 6.11.2 Function elements........................68 6.11.3 Adjusting settings in the quick access menu................71 6.12 Messages and audio feedback......................71 6.12.1 Messages..........................71 6.12.2 Wizard............................73 6.12.3 Audio feedback.........................73 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 6 7 .7 .2 Adding a startup screen......................114 7 .7 .3 Configuring the unit for screenshots..................116 Backing up data..........................117 7 .8.1 Back up settings........................117 7 .8.2 Back up user files........................118 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 7 Acquiring the minimum, maximum, and range..............163 9.2.7 Sending measured values to a computer................164 9.2.8 Displaying the diameter......................165 9.2.9 Performing a relative measurement..................166 9.2.10 Measure with dial gage......................167 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 8 Contents 10 File management..........................169 10.1 Overview............................. 170 10.2 File types.............................171 10.3 Managing folders and files....................... 171 10.4 Viewing and opening files........................ 173 10.5 Exporting files............................ 174 10.6 Importing files............................ 175 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 9 Nonlinear error compensation (NLEC)................... 194 11.6.5 Squareness error compensation (SEC)..................195 11.6.6 Alias assignment for axis names...................195 11.6.7 <Axis name> (settings of the axis)..................196 11.6.8 Encoder..........................197 11.6.9 Reference marks (Encoder)....................201 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 10 12.8 Reset to shipping conditions......................219 13 What to do if ..........................221 13.1 Overview............................. 222 13.2 System or power failure........................222 13.2.1 Restoring the firmware......................222 13.2.2 Restore settings........................223 13.3 Malfunctions............................223 13.3.1 Troubleshooting........................223 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 11 Product dimensions with Single-Pos stand................231 15.3.2 Product dimensions with Duo-Pos stand................231 15.3.3 Product dimensions with Multi-Pos stand................232 15.3.4 Product dimensions with Multi-Pos holder................232 16 Index...............................233 17 List of figures..........................235 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 13: Fundamentals

    Fundamentals...
  • Page 14: Overview

    You can download the current version of the software here: www.heidenhain.de To download the installation file from the HEIDENHAIN Portal, you need access rights to the Software portal folder in the directory of the appropriate product.
  • Page 15: Documentation On The Product

    If the ID numbers and indexes as well as the firmware versions match, the documentation is valid If the part numbers and indexes do not match so that the documentation is not valid, you will find the current documentation for the product at www.heidenhain.de. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 16: Notes On Reading The Documentation

    Have you found any errors or would you like to suggest changes? We continuously strive to improve our documentation for you. Please help us by sending your suggestions to the following e-mail address: userdoc@heidenhain.de HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 17: Storage And Distribution Of The Documentation

    These instructions must be read and observed by every person who performs any of the following tasks: Mounting Installation Commissioning and configuration Operation Service, cleaning and maintenance Troubleshooting Removal and disposal HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 18: Target Groups According To User Types

    ... items supplied with the product ... accessories for the product ✓ ✓ ... correct mounting of the product 4 "Mounting" ✓ ✓ ... correct installation of the product 5 "Installation" HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 19 ... the technical data of the product ✓ ✓ ✓ 15 "Specifications" ... product dimensions and mating dimensions (drawings) This chapter enables accessing the content of these ✓ ✓ ✓ 16 "Index" instructions according to specific topics. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 20: Notes In This Documentation

    The behavior of the functions depends on the configurable machine settings The book symbol represents a cross reference to external documentation, e.g. the documentation of your machine tool builder or other supplier. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 21: Symbols And Fonts Used For Marking Text

    Identifies an item of a list Example: TTL interface EnDat interface Bold Identifies menus, displays and buttons Example: Tap Shut down The operating system shuts down Turn the power switch off HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 23: Safety

    Safety...
  • Page 24: Overview

    The information on their intended use, which is provided in the respective documentation, must be observed. Improper use In particular, the products of the GAGE-CHEK 2000 series must not be used in the following applications: Use and storage outside the operating conditions specified in "Specifications"...
  • Page 25: Personnel Qualification

    Provide all materials and means necessary in order for the personnel to complete the assigned tasks Ensure that the device is operated only when in perfect technical condition Ensure that the device is protected from unauthorized use HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 26: General Safety Precautions

    Functional ground connection as per IEC/EN 60204-1. Observe the information on installation. Product seal. Breaking or removing the product seal will result in forfeiture of warranty and guarantee. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 27: Electrical Safety Precautions

    Damage to internal parts of the product! If you open the product, the warranty and the guarantee will become void. Never open the housing Only the product manufacturer is permitted to access the inside of the product HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 29: Transport And Storage

    Transport and storage...
  • Page 30: Overview

    PDF issue of the Operating Instructions on a memory medium in the currently available languages Product Evaluation Unit GAGE-CHEK 2000 Installation Instructions Printed issue of the Installation Instructions in the currently available languages HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 31: Accessories

    (for setting presets) Power cable Power cable with European 223775-01 plug (type F), length: 3 m RS-232 connecting RS-232 connecting cable, 366964-xx cable complete with two 9-pin D- sub connectors (female) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 32 STEINWALD datentechnik GmbH to the product, the data interface is configured automatically and is ready immediately. The data format is used for measured value output. The settings are not Steinwald configurable. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 33: In Case Of Damage In Transit

    Storage of the product Package the product as described above Observe the specified ambient conditions Further information: "Specifications", Page 227 Inspect the product for damage after any transport or longer storage times HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 35: Mounting

    Mounting...
  • Page 36: Overview

    Allen key, size 2.5 (Duo-Pos stand) Materials for mounting on a supporting surface The unit must be mounted to a stand or a holder to ensure the correct and intended use of the product. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 37: Mounting On Single-Pos Stand

    Route the cable from behind through the opening of the stand and then through the lateral openings to the connections Figure 2: Product mounted on Single-Pos Figure 3: Cable routing on Single-Pos stand stand "Product dimensions with Single-Pos stand", Page 231 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 38: Mounting On Duo-Pos Stand

    Route the cable from behind through the two supports of the stand and then through the lateral openings to the connections Figure 4: Product mounted on Duo-Pos stand Figure 5: Cable routing on Duo-Pos stand "Product dimensions with Duo-Pos stand", Page 231 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 39: Mounting On Multi-Pos Stand

    Route the cable from behind through the two supports of the stand and then through the lateral openings to the connections Figure 6: Product mounted on Multi-Pos stand Figure 7: Cable routing on Multi-Pos stand Further information: "Product dimensions with Multi-Pos stand", Page 232 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 40: Mounting On Multi-Pos Holder

    Route the cable from behind through the two supports of the holder and then through the lateral openings to the connections Figure 8: Product mounted on Multi-Pos Figure 9: Cable routing on Multi-Pos holder holder Further information: "Product dimensions with Multi-Pos holder", Page 232 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 41: Installation

    Installation...
  • Page 42: Overview

    (ESD). It is essential to observe the safety precautions for handling ESD-sensitive components Never touch connector pins without ensuring proper grounding Wear a grounded ESD wristband when handling device connections HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 43: Device Overview

    Remove dust protection caps only when connecting measuring devices or peripherals If you remove a measuring device or peripheral, re-attach the dust protection cap to the connection The type of connections for encoders may vary depending on the product version. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 44: Figure 10: Rear Panel On Devices With Id 1089181-01

    X116: RJ45 Ethernet connection for communication and data exchange with subse- quent systems or PC X113: 15-pin D sub connection for touch probes (e.g., HEIDENHAIN touch probe) X100: Power switch and power connection HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 45: Connecting Encoders

    Pin layout of X1, X2, X3 11 µA EnDat 2.2 R– 11 μA Inter- EnDat DATA shield CLOCK A– Sense B– Sense 11 μA 1– 2– DATA CLOCK EnDat Pin layout of X21, X22, X23 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 46: Connecting Touch Probes

    B – Probe signals, readiness TP – Touch Probe, normally closed Digital inputs: Digital outputs: X113 X113 Din 0...3 DC 5 V DC 5 V Dout 0 Touch probe: X113 DC 5 V HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 47: Wiring Switching Inputs And Outputs

    "Device overview", Page 43 If the cable connectors include mounting screws, do not overtighten them The digital or analog inputs and outputs must be assigned in the device settings of the respective switching function. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 48: Connecting Input Devices

    Connect USB mouse or USB keyboard to USB Type-A port (X32). Make sure the USB cable connector is fully inserted Further information: "Device overview", Page 43 Pin layout of X32, DC 5 V Data (–) Data (+) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 49: Connecting A Network Peripheral

    Comply with the pin layout Connect the power connection to a 3-wire grounded power outlet using a power cable that meets requirements Further information: "Device overview", Page 43 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 50 Installation | Connecting the line voltage Pin layout X100 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 51: Basic Operation

    Basic operation...
  • Page 52: Overview

    Operation using the mouse The following overview describes the different gestures for operating the touchscreen or the mouse: Tapping Means touching the screen briefly with your fingertip Means pressing the left mouse button once HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 53 Single pressing and holding down of the left mouse button while simultaneously moving the mouse; the starting point and end point of the movement are not clearly defined The actions initiated by swiping include Changes views HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 54: General Operating Elements And Functions

    Tap + or – until the desired value is displayed Long-press + or – to scroll through the values more quickly The selected value is displayed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 55 Processes that have already been concluded cannot be undone. Tap Undo The last action is undone To add a feature, tap Add The new feature is added Close Tap Close to close a dialog HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 56: Gage-Chek 2000 - Switch-On And Switch-Off

    Deactivating energy-saving mode Tap anywhere on the touchscreen An arrow appears at the bottom of the screen Drag the arrow up The screen is switched on and shows the user interface last displayed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 57: Switching Off The Gage-Chek 2000

    Before a new user can log in, the logged-in user has to log out. The product provides various authorization levels that grant the user full or restricted access to management and operation functionality. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 58: User Login

    Tap Log out The user is logged out All functions of the main menu are inactive, except for Switch The product can only be used again after a user has logged in HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 59: Setting The Language

    The Reference symbol stops blinking upon successful completion of the reference mark search "Operating elements of the position display", Page 66 Further information: Further information: "Activating the reference mark search", Page 113 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 60: User Interface

    Inspector after starting up. Further information: "Measure menu", Page 62 If automatic user login is not activated, then the product opens the User login menu. Further information: "User login menu", Page 64 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 61: Main Menu Of The User Interface

    OEM user type) is logged in, then the gear symbols appears. Settings Settings of the product, such as setting up users, configuring sensors, or updating the firmware Further information: "Settings menu ", Page 65 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 62: Measure Menu

    The user interface for measuring and positioning is displayed Short description Figure 14: Measure menu The workspace shows the current position of the measuring plate The Inspector contains the quick access menu and the function elements HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 63: File Management Menu

    The USB mass storage products and the network drives are displayed with their name or drive designation. Further information: "File management", Page 169 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 64: User Login Menu

    The User login menu shows the logged-in user in the column on the left. The login of a new user is displayed in the right-hand column. To log in another user, the logged-in user must first log out. "User login and logout", Page 57 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 65: Settings Menu

    Further information: "Settings", Page 177 The product provides various authorization levels that grant the user full or restricted access to management and operation functionality. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 66: Switch-Off Menu

    Holding down the axis key: sets the current position as zero point Reference mark search performed successfully Reference mark search not performed or no reference mark detected Minimum: Lowest measured value (if the MinMax function is active) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 67: Customizing The Workspace

    To display the Inspector, tap the tab 6.11 Using the Inspector The Inspector is only available in the Measure menu. Activation Tap Measure in the main menu The user interface for measuring, constructing and defining is displayed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 68: Controls Of The Inspector

    Send measured values to computer; tapping activates the automatic output of measured values according to the configuration; the data is transferred when the stylus is deflected "Sending measured values to a Further information: computer", Page 164 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 69 Saving of measured values of a reference part as master or adopting position values of the position display as master; corresponding axes can be selected "Configure the Mastering function", Further information: Page 134 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 70 Drag the function element to the right Tap Open The Open configuration dialog is opened Navigate to the folder in which the file is saved Tap the desired XMG file Tap Open The file is opened HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 71: Adjusting Settings In The Quick Access Menu

    The messages are displayed upon occurrence of the message cause or via tapping on the Messages display area at the top left of the screen. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 72 The message disappears To close a message that potentially has an effect on the application, tap OK If applicable, the message will now be taken into account by the application The message disappears HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 73: Wizard

    The available sounds are grouped into categories. The sounds differ within a category. You can define the audio feedback settings in the Settings menu. "Sounds", Page 182 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 75: Commissioning

    Commissioning...
  • Page 76: Overview

    If the password is no longer known, contact a HEIDENHAIN service agency. Confirm the entry with RET Tap Log in The user is logged in The product opens the Measure menu HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 77: Performing The Reference Mark Search After Startup

    The language selected for the user is indicated by a national flag in the Language drop-down list Select the flag for the desired language from the Language drop-down list The user interface is displayed in the selected language HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 78: Changing The Password

    You are logged on as a user of the OEM type (see "Logging in for commissioning", Page 76). Basic settings Activating the Software options Setting the date and time Setting the units of measure Configuring a touch probe Configuring a touch probe HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 79: Basic Settings

    "Checking the Software options", Page 81 Requesting license key You can request a license key by using the following procedure: Reading out device information for the license key request Creating a license key request HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 80 Further information: "Moving a file", Page 172 Contact a HEIDENHAIN service agency and submit the file you created in order to request a license key for the product The license key and the license file are generated and sent by...
  • Page 81 On the overview page, you can check which Software options are enabled for the product. Tap Settings in the main menu Tap Service Open in the sequence Software options Overview A list of enabled Software options is displayed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 82: Setting The Date And Time

    To set the rounding method, tap the corresponding drop- down list and select the rounding method To set the number of decimal places displayed, tap – or + Further information: "Units", Page 183 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 83: Configuring A Touch Probe

    11 µA or TTL interface: The parameters must be configured manually For the parameters of HEIDENHAIN encoders that are typically connected to the product, refer to the overview of typical encoders. Further information: "Overview of typical encoders", Page 85...
  • Page 84: Configure Alias Assignment For Axis Names

    MyFormat1.xml or another format file you created by adding the axis names that you assigned. Further information: "Creating your own data format", Page 146 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 85: Overview Of Typical Encoders

    Commissioning | Configuring the axes 7.6.2 Overview of typical encoders The following overview lists the parameters of the HEIDENHAIN encoders that are typically connected to the product. When connecting other encoders, refer to the encoder's documentation for the required parameters.
  • Page 86 0.4 μm or 4 μm with 5-fold evaluation Length Interface Measuring step Maximum gaugeseries traverse path AT 121x EnDat 2.2 23 nm 12 mm AT 301x EnDat 2.2 368 nm 30 mm HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 87: Configuring Axes For Encoders With Endat Interface

    An encoder with EnDat interface is connected to the product. The configuration procedure is the same for each axis. The procedure will now be explained using one axis as an example. Tap Settings in the main menu HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 88 To view the electronic ID label of the encoder, tap ID label In order to see the results of the encoder diagnosis, tap Diagnosis "<Axis name> (settings of the axis)", Page 196 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 89 One: The encoder has one reference mark Coded: The encoder has distance-coded reference marks If the linear encoder has coded reference marks, enter the Maximum traverse path (see " Linear encoders", Page 85) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 90 Positive: The direction of traverse is in the counting direction of the encoder Negative: The direction of traverse is opposite to the counting direction of the encoder Further information: "<Axis name> (settings of the axis)", Page 196 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 91 The ascertained line count remains stored if you select a different display mode after the teach sequence. "Settings for encoders with interfaces of the type 1 V Further information: 11 A ", Page 197 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 92: Configuring Axes For Encoders With Ttl Interface

    Maximum traverse path (see " Linear encoders", Page 85) If the angle encoder has coded reference marks, enter the parameter for the Nominal increment(see "Angle encoders", Page 87) Confirm the entry with RET HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 93 Positive: The direction of traverse is in the counting direction of the encoder Negative: The direction of traverse is opposite to the counting direction of the encoder "<Axis name> (settings of the axis)", Page 196 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 94: Performing Error Compensation

    Segmented linear error compensation (SLEC): The axis is divided into multiple segments with the help of a maximum of 200 supporting points. A distinct compensation factor is defined and applied for every segment. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 95 Enter the length or angle of the actual traverse path determined by measuring (actual length or actual angle) Confirm the entry with RET Activate Compensation with the ON/OFF slide switch Further information: "Linear error compensation (LEC)", Page 203 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 96 To switch to the previous display, tap Back twice Activate Compensation with the ON/OFF slide switch The error compensation for the axis is applied Further information: "Segmented linear error compensation (SLEC)", Page 203 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 97 The following options are available for measuring the nominal and actual values of the supporting points: Acquiring nominal values Read deviations from the calibration standard (ACF) Create a table of supporting points manually HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 98 Open in succession: Axes General settings Error compensation Nonlinear error compensation (NLEC) Use the ON/OFF slider to deactivate the Compensation function The table of supporting points is now released for editing HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 99 Enter the values, separated by tab stops as shown in the schema below The following specifications apply to import files: File names must not contain diacritics or special characters Use a point as decimal character HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 100 20.0003 -0.00010 39.9998 24.9981 39.9979 49.9999 40.0001 75.0004 40.0021 100.0019 40.0008 0.00003 59.9992 25.0000 60.0018 50.0001 60.0003 75.0020 59.9990 100.0001 60.0001 -0.00003 80.0021 24.9979 80.0004 50.0020 79.9991 75.0001 79.9985 100.0010 80.0002 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 101 The file contains an additional row with X and Y values for each supporting point. The supporting points can be indicated row by row or in a meander- shaped sequence. The product adapts the reading direction automatically. Row-by-row reading direction Meander-shaped reading direction HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 102 XML: The actual values are imported from the file TXT: The actual values are imported from the file Extended TXT: The actual values are corrected by the deviations HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 103 20.0020 25.0013 20.0021 50.0013 20.0022 75.0005 20.0023 99.9996 20.0003 -0.00010 39.9998 24.9981 39.9979 49.9999 40.0001 75.0004 40.0021 100.0019 40.0008 0.00003 59.9992 25.0000 60.0018 50.0001 60.0003 75.0020 59.9990 100.0001 60.0001 -0.00003 80.0021 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 104 Actual value of the second supporting point on the X supporting point on the Y axis axis The file contains an additional row with X and Y values for each supporting point. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 105 Station (5, 2) -0.00040 0.00030 Station (1, 3) -0.00010 -0.00020 Station (2, 3) -0.00190 -0.00210 Station (3, 3) -0.00010 0.00010 Station (4, 3) 0.00040 0.00210 Station (5, 3) 0.00190 0.00080 Station (1, 4) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 106 Deviation of the second supporting point on the X and Y axes The file contains a Station (x, y) section for each supporting point, indicating the deviations on the X and Y axes. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 107 The following example shows a grid of 5 x 5 supporting points with a spacing of 25 mm on the X axis and 20 mm on the Y axis. Example <?xml version="1.0" encoding="UTF-8"?> <configuration> <base id="Settings"> <group id="CellSize"> <element id="x">25</element> <element id="y">20</element> </group> <group id="General"> <element id="enabled">false</element> HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 108 <element id="3-3" Y="59.999000000000002" X="75.001999999999995"/> <element id="4-3" Y="60.000100000000003" X="100.0001"/> <element id="0-4" Y="80.002099999999999" X="-3.0000000000000001"/> <element id="3-4" Y="79.998500000000007" X="75.000100000000003"/> <element id="2-4" Y="79.999099999999999" X="50.002000000000002"/> <element id="4-4" Y="80.000200000000007" X="100.001"/> </group> </base> <base id="version" build="0" minor="4" major="1"/> </configuration> HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 109 Actual value of the supporting point in X="25.001200000000001"/> mm. In this example: X = -0.001 Y = 25.001200000000001 For each supporting point, the group contains an additional element with these parameters. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 110 Tap Settings in the main menu Open in succession: Axes General settings Error compensation Nonlinear error compensation (NLEC) Tap Table of supporting points Enter the actual values of the supporting points Confirm each entry with RET HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 111 Enter the measured values of the reference standard (= nominal values) Activate Compensation with the ON/OFF slide switch Squareness error compensation is applied from the next measurement "Squareness error compensation (SEC)", Page 195 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 112: Coupling Axes

    -: principal axis - coupled axis The position values of both axes are offset against each other according to the selected type of calculation Further information: "<Axis name> (settings of the axis)", Page 196 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 113: Activating The Reference Mark Search

    Documentation: Adding the OEM documentation, e.g. service information Startup screen: Defining a startup screen with the OEM's company logo Screenshots: Configuring the unit for screenshots with the program Screen- shotClient HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 114: Adding Documentation

    File type: PNG or JPG Resolution: 96 ppi Image format: 16:10 (other formats will be scaled proportionally) Image size: Max. 1280 x 800 px HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 115 Disconnect the USB mass storage device When you save the user files, the OEM-specific opening screen is also saved and can be restored. Further information: "Back up user files", Page 118 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 116: Configuring The Unit For Screenshots

    For a detailed description, please refer to the GAGE-CHEK 2000 Demo User's Manual. This User's Manual is available in the "Documentation" folder of the product website. "Demo software for the product", Page 14 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 117: Backing Up Data

    Confirm the entry with RET Tap Save as Tap OK to confirm the successful backup of the configuration The configuration file was backed up "Back up and restore", Page 206 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 118: Back Up User Files

    Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 119 Setup...
  • Page 120 Setup user or OEM user for the assigned password. If the password is no longer known, contact a HEIDENHAIN service agency. Confirm the entry with RET Tap Log in HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 121 The language selected for the user is indicated by a national flag in the Language drop-down list Select the flag for the desired language from the Language drop-down list The user interface is displayed in the selected language HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 122 Confirm entry with RET Enter the new password and repeat it Confirm entry with RET Tap OK Close the message with OK The new password is available the next time the user logs in HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 123 If the product is disconnected from the power source while it is on, the configuration data can be lost or corrupted. Back up the configuration data and keep the backup for recovery purposes HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 124 To set the number of decimal places displayed, tap – or + "Units", Page 183 Further information: Entering and configuring users The following user types, which have different rights, are defined in the product's factory default settings: Setup Operator HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 125 Use the ON/OFF sliding switch to show or hide the contents Tap OK A message appears Close the message with OK The user is created with the basic data. The user can then further edit the data himself later HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 126 Close the message with OK To change the language, select the flag for the desired language in the Language drop-down list Use the ON/OFF slider to activate or deactivate the Auto login function HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 127 The Operating Instructions are copied to the product Any existing Operating Instructions will be overwritten Confirm the successful transfer with OK The Operating Instructions can be opened and displayed on the product HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 128 Confirm the entires with RET Enter the Preferred DNS server and, if required, the Alternative DNS server Confirm the entires with RET The configuration of the network connection is applied "Network", Page 186 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 129 "Connecting input devices", Page 48 Tap Settings in the main menu Tap General Tap Input devices Select the desired option from the Mouse substitute for multitouch gestures drop-down list "Input devices", Page 181 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 130 Measure menu. The procedure is identical for all function elements. For more information on the function elements, please refer to the “Basic operation” chapter. Further information: "Function elements", Page 68 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 131 Determines the coordinate of the traversed axis Determining the circle center Determine the center point of a circular form (hole or cylinder) (four probing procedures) Determines the coordinates of both axes and the circle diameter HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 132 The following steps are necessary in order to configure the MinMax function: Add the MinMax function element to the Inspector Select measured values that are to be captured and displayed when the MinMax function is active HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 133 Drag an empty field of the Inspector to the left into the workspace A dialog box for selecting the function element appears Tap the D/R function element Tap Close The D/R function element is available in the Inspector HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 134 The following steps are necessary in order to configure the Mastering function: Add the Mastering function element to the Inspector Selecting axes Enter or apply values that should be set HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 135 Zero the current axis values of your axis or use Mastering to reset your axis. "Configure the Mastering function", Page 134 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 136: Figure 21: Example Of An Absolute Measurement

    Drag an empty field of the Inspector to the left into the workspace The dialog for selecting the function element is opened Tap the dial gage function element Tap Close The dial gage function element is available in the Inspector HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 137: Figure 23: Example Of A Value Input For An Axis

    After you activated the dial gage for your axis, you can enter values. The device offers two possibilities to enter values: Enter values individually Enter values symmetrically Figure 23: Example of a value input for an axis HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 138 You can access the preset table on the status bar. The preset table contains the absolute positions of the presets with respect to the reference mark. You can save a maximum of 99 datums in the datum table. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 139 To protect the entry for a preset from accidental changes or deletion, tap the Lock symbol next to the preset's entry The symbol changes and the entry is locked Tap Close The Preset table dialog box is closed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 140 In order to apply a measured position value as the new zero point, leave the input fields under Set position values empty Tap Confirm in the wizard The position is applied as preset HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 141 STEINWALD datentechnik GmbH to the product, the data interface is configured automatically and is ready immediately. The data format Steinwald is used for measured value output. The settings are not configurable. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 142 As an alternative to the functions of the Inspector described here, you can use the Trigger measured value output switching function for the measured-value output. Further information: "Switching functions", Page 190 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 143 Standard, Steinwald, MyFormat1, and all self-created data formats. In order to assign a data format to a function, select the desired data format in the respective drop-down list Further information: "Data transfer", Page 189 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 144 MyFormat1.xml or another format file you created by adding the axis names that you assigned. Further information: "Creating your own data format", Page 146 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 145: Figure 24: Example Of A Transfer For The X And Y Axes With The Active Minmax Function In The Data Format

    Algebraic sign (< 0, then minus sign) Digits before the decimal point Decimal point Digits after the decimal point Backwards movement of the cursor to the beginning of the line (Carriage return) Line feed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 146: Figure 25: Example Of A Transfer For The X And Y Axes With The Active Minmax Function In The Data Format

    Then you can copy the new file to the file storage area of the product and assign it to a function. Data formats are saved as XML files HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 147: Figure 26: Data Format Myformat1.Xml

    Figure 26: Data format MyFormat1.xml Header Name of the data format that appears in the menu Settings General data format settings Settings of the axes Footer End of the data format HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 148 (current) value receives the current time upon sending. The second (previous) value retains its original time stamp element "true" A page break is inserted after the time stamp newlineAfterTimestamp Only if attribute writeTimestamp="true" HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 149 "0" Number of digits before the decimal separator to which the measured value is rounded according to commercial practice Example: Measured value: 43.67 digits="4" Measured value output: 0043.67 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 150 Tap one of the following function elements: Manual output of measured values Touch-probe triggered output of measured values Continuous output of measured values Tap Close The function element is available in the Inspector HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 151: Figure 27: Representation Of Selected Contents For Measured Value Output

    If measured values are missing then the value 0 is sent. Further information: "Configuring the MinMax function", Page 132 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 152 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 153 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 155 Measuring...
  • Page 156 Certain influences, such as incident light, also have a negative effect on the measurement uncertainty. Eliminate or avoid environmental influences as far as possible HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 157 Open in the sequence Axes General settings Reference marks Tap Start The Reference symbol blinks Follow the instructions of the wizard The Reference symbol stops blinking upon successful completion of the reference mark search HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 158 The position value is applied to the current position The entered position value is linked with the current position and overwrites the active preset in the preset table The active preset is applied as the new value HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 159 In order to apply a measured position value as the new zero point, leave the input fields under Set position values empty Tap Confirm in the wizard The position is applied as preset HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 160: Figure 28: Measure Menu

    Move to the desired position or acquire measured values The result appears in the position display You can send the measured values to a computer "Sending measured values to a computer", Page 164 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 161: Figure 29: Measure Menu With Touch Probe

    "Configuring a touch probe", Page 83 Further information: Further information: "Configuring probing functions", Page 131 Further information: "Configuring the axes", Page 83 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 162: Figure 30: Inspector Showing The Function Elements For The Probing Functions

    If the Select preset function is activated, a dialog appears in which you can apply the position as the new preset If the MVO function is enabled, the product transmits the measured values to the computer HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 163: Figure 31: Measure Menu With Active Minmax Function

    The MinMax function is paused, and the icon turns gray Tap the gray MinMax function element in order to continue capturing To complete the capture processes, tap the green MinMax function element HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 164 In order to deactivate the function, tap the Continuous output of measured values function element again In addition, you can activate automatic data transfer individually for each probing function. Further information: "Configuring probing functions", Page 131 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 165: Figure 32: Measure Menu With Activated D/R Function

    For the axes whose position value is converted, the diameter symbol appears In order to display the radius, tap the D/R function element again This conversion is deactivated for all axes HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 166: Figure 33: Measure Menu With Activated Relative Function

    Perform the desired measurement In order to end the relative measurement, tap the Relative function element again The position display switches over to the standard view The preset table is enabled for editing HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 167: Figure 34: Overview

    The measured value exceeds the warning limit but still lies within the tolerance limit. The measured value exceeds the tolerance limit. Opening the overview To open the overview: In the Inspector, tap the function element dial gage The overview opens HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 168: Figure 35: Single View Of The Dial Gage

    To switch from the overview to the single view of an axis, proceed as follows: Tap the desired single view, Swipe across the touchscreen from right to left until the desired single view appears The single view opens HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 169 File management...
  • Page 170: Figure 36: File Management Menu

    Tap File management in the main menu The file management user interface is displayed Figure 36: File management menu List of available storage locations List of folders in the selected storage location HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 171 Drag the icon of the folder you want to move to the right The operating elements are displayed Tap Move to In the dialog, select the folder to which you want to move the folder Tap Select The folder is moved HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 172 Tap Select The file is copied If you copy a file to the folder it is stored in, the suffix "_1" is appended to the name of the copied file. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 173: Figure 37: File Management Menu With Preview Image And File Information

    A preview image (only with PDF and image files) as well as information about the file are displayed Figure 37: File management menu with preview image and file information Tap View The file contents are displayed Tap Close to close the view HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 174 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 175 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 177 Settings...
  • Page 178 Configuration of interfaces and network drives User Configuration of users Axes Configuration of connected encoders and error compensation Service Configuration of software options, service functions and information Activation Tap Settings in the main menu HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 179 <Axis name> (settings of the axis) 11.6.6 "Alias assignment for axis names" 11.7.1 "Firmware information" 11.7.2 "Back up and restore" 11.7.3 "Firmware update" Service 11.7.4 "Reset" 11.7.5 "OEM area" 11.7.7 "Documentation" 11.7.8 "Software options" HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 180 Tap and drag: Touch the touchscreen and drag the arrow upwards from the lower edge Tap: Touch the touchscreen Tap or axis movement: Touch the touchscreen or move the axis Default setting: Tap and drag HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 181 Off (only multitouch): Operation only possible with the touchscreen, the mouse is deactivated Default setting: Auto (until first multitouch) USB keyboard layout If a USB keyboard is connected: Language selection of the keyboard assignment HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 182 Default setting: Current system time Date format Format in which the date is displayed Settings: MM-DD-YYYY: month, day, year DD-MM-YYYY: day, month, year YYYY-MM-DD: year, month, day Default setting: YYYY-MM-DD (e.g. "2016-01-31") HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 183 Round to 0 and 5: Decimal digits ≤ 24 or ≥ 75 are rounded to 0, decimal digits ≥ 25 or ≤ 74 are rounded to 5 Default setting: Commercial HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 184 Meaning and function Operating Instructions Display of the operating instructions stored on the product Default: no document; the document in the desired language can be added "Documentation", Page 208 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 185 Setting range: ≥ 0.0001 Default value: 6.0000 Evaluation of the ready signal Possibility of setting whether the ready signal of the touch probe should be evaluated Setting range: ON or OFF Default value: ON HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 186 IPv6 standard gateway Network address of the router connecting a network Preferred DNS server Primary server for mapping the IP address Alternative DNS server Optional server for mapping the IP address HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 187 Kerberos V5 authentication and packet signing NTLM password hashing NTLM password hashing with signing NTLMv2 password hashing NTLMv2 password hashing with signing Default value: None Configuration of the Mount options Settings: Default value: nounix,noserverino HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 188 Selection of the parity bit for verification Settings: None Even Space Mark Stop bits Selection of the stop bit for synchronization Settings: 1 bit 2 bits Flow control Selection of the data flow Settings: None Hardware Xon/Xoff HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 189 Data format for continuous data Selection of the data format for measured value output transfer Settings: Standard Steinwald MyFormat1 (template for copying) Any data formats you have defined yourself Default value: Standard HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 190 The position-dependent switching functions enable you to set logical outputs depending on the position of an axis in a defined reference system. Parameter Explanation Output Selecting the desired output X113.04 (Dout 0) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 191 Default value: Setup – First name First name of the user Default value: – – Department Department of the user Default value: – – Group Group of the user Default value: setup HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 192 Settings ► User ► + Path: Parameter Explanation Adds a new user of the type Operator Further information: "Entering and configuring users", Page 124 It is not possible to add further OEM and Setup-type users. HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 193 Start starts the reference mark search and opens the workspace Reference mark search status Indicates whether the reference mark search was successful Display: Successful Unsuccessful Stop of reference mark search Indicates whether the reference mark search was canceled Display: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 194 Number of supporting points Number of measuring points for error compensation on both axes (X and Y) of the encoder Setting range: 1 ... 99 (X and Y) Default value: 2 (X and Y) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 195 Parameter Explanation Setting range: 00 ... 99 and aA ... xX Default value: X (for C1) Default value: Y (for C2) Default value: Z (for C3) HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 196 Calculation type for the position values of the principal (main) axis and coupled axis Settings: +: The position values are added (principal axis + coupled axis) -: The position values are subtracted (principal axis – coupled axis) Default value: + HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 197 Settings for encoders with interfaces of the type 1 V Parameter Explanation Encoder input Assignment of the encoder input to the axis of the product Settings: Not connected Further information: "Device overview", Page 43 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 198 Analog filter frequency Frequency value of the analog low-pass filter Settings: 33 kHz: Suppression of interference frequencies above 33 kHz 400 kHz: Suppression of interference frequencies above 400 kHz Default value: 400 kHz HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 199 Settings for encoders with interfaces of the TTL type Parameter Explanation Encoder input Assignment of the encoder input to the axis of the product Settings: Further information: "Device overview", Page 43 Interface Automatically detected TTL interface type HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 200 A warning or error message is displayed if one of the limit values for error monitoring is exceeded. The limit values depend on the signal of the connected encoder: Error message with frequency ≥ 5 MHz HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 201 For encoders with TTL interface: Interpolation value of the encoders and integrated interpolation for the evaluation of the coded reference marks. Settings: None 2-fold 5-fold 10-fold 20-fold 50-fold Default value: None HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 202 Current position for reference Apply applies the current position as an offset (in mm or point shift degrees, depending on the selected encoder model) between the reference mark and the zero point HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 203 Opens the table of supporting points for manual editing Create table of supporting points Opens the menu for creating a new Table of supporting points Further information: "Create table of supporting points", Page 204 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 204 The start point defines the position starting from which the compensation is applied to the axis Default value: 0.00000 Create Creates a new table of supporting points based on the entries HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 205 Version number of the programming module system.xml Version number of the system parameters axes.xml Version number of the axis parameters encoders.xml Version number of the encoder parameters ncParam.xml Version number of the NC parameters HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 206 Restoring of user files of the product Further information: "Restore user files", Page 217 Back up user files Backing up of user files of the product Further information: "Back up user files", Page 118 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 207 Reset to shipping conditions Resetting of the settings to the factory default setting and deletion of the user files from the unit’s memory area "Reset to shipping conditions", Page 219 Further information: HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 208 USB mass storage device to the product. The latest version can be downloaded from the download area at www.heidenhain.de. Parameters Explanation Add Operating Instructions Adding the Operating Instructions in the desired language HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 209 Activation of the software options via license key or license file "Activating a license key", Page 80 Further information: Reset trial options Reset of the trial options by entering a license key HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 211 Servicing and maintenance...
  • Page 212 To deactivate the cleaning mode, tap anywhere on the touchscreen An arrow appears at the bottom of the screen Drag the arrow up The screen is switched on and shows the user interface last displayed HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 213 Operating defective devices may result in serious consequential damage. Do not repair or operate the device if it is damaged Replace defective devices immediately or contact a HEIDENHAIN service agency The following steps are only to be performed by electrical specialists.
  • Page 214 Starting a firmware update Tap Settings in the main menu Tap Service Open in the sequence Firmware update Continue The service application is launched HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 215 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 216 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 217 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 218 Tap File management in the main menu Navigate to the list of storage locations Tap Safely remove The message "The storage medium can be removed now." appears Disconnect the USB mass storage device HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 219 Tap OK to confirm shutdown of the device The product is shut down All settings are reset and the user files are deleted To restart the product, switch it off and then back on HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 221 What to do if ...
  • Page 222 13.2.1 Restoring the firmware On a computer, create the folder "heidenhain" on a USB mass storage device (FAT32 format). In the "heidenhain" folder, create the folder "update" Copy the new firmware to the "update" folder Rename the firmware "recovery.dro"...
  • Page 223 Qualified movement of the encoder of the encoder personnel Contact the encoder manufacturer's service agency Axes are miscounting Incorrect settings of Check the encoder settings Qualified the encoder Page 89 personnel HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 224 To reset the OEM password, contact the HEIDENHAIN service agency Data transfer does not Incorrect settings for Check the configuration of Qualified work data transfer the interface in the settings personnel HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 225 Removal and disposal...
  • Page 226 Forward the product and the backup battery to recycling in accordance with the applicable local disposal regulations If you have any questions about the disposal of the product, please contact a HEIDENHAIN service agency HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 227 Specifications...
  • Page 228 Four digital inputs: TTL DC 0 V ... +5 V One digital output TTL DC 0 V ... +5 V Maximum load 1 k Ω Max. cable length with HEIDENHAIN cable 30 m HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 229 Front panel and side panels: IP 65 Rear panel: IP 40 Mass 1.3 kg With Single-Pos stand: 1.35 kg With Duo-Pos stand 1.45 kg With Multi-Pos stand: 1.95 kg With Multi-Pos holder: 1.65 kg HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 230: Figure 38: Housing Dimensions For Products

    Figure 38: Housing dimensions for products HEIDENHAIN X113 X100 AC 100 V ... 240 V 50 Hz ... 60 Hz (max. 38 W) 37 .5 15.5 X116 67 .5 Figure 39: Dimensions of the rear panel HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 231: Figure 40: Product Dimensions With Single-Pos Stand

    15.3.2 Product dimensions with Duo-Pos stand ISO 7092-4 ISO 7380-M4x8 ISO 7092-4 Md = 2.6 Nm ISO 7380-M4x8 Md = 2.6 Nm 132±2 150±2 Figure 41: Product dimensions with Duo-Pos stand HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 232: Figure 42: Product Dimensions With Multi-Pos Stand

    Figure 42: Product dimensions with Multi-Pos stand 15.3.4 Product dimensions with Multi-Pos holder ISO 14581-M4x8 Md = 2.6 Nm Md = 15 Nm max. Figure 43: Product dimensions with Multi-Pos holder HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 233 Firmware update...... 214 Dial gage........135 Data format, selecting..143 Folder Activate axes....... 137 Data format parameters..144 copying........ 172 Activate the switching Data formats......146 creating........ 171 function........ 138 Functions......142 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 234 Segmented linear error customize......67 slide switch......55 compensation (SLEC)....96 toggle switch......55 Settings undo........55 Backing up....117, 152 Operating Instructions....16 Menu........65 updating....... 127 Restore........ 218 Operation HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 235 Dimensions of the rear panel....................230 Figure 40: Product dimensions with Single-Pos stand................231 Figure 41: Product dimensions with Duo-Pos stand................231 Figure 42: Product dimensions with Multi-Pos stand................232 Figure 43: Product dimensions with Multi-Pos holder................232 HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020...
  • Page 236 DR. JOHANNES HEIDENHAIN GmbH Dr.-Johannes-Heidenhain-Straße 5 83301 Traunreut, Germany  +49 8669 31-0  +49 8669 32-5061 E-mail: info@heidenhain.de Technical support  +49 8669 32-1000 Measuring systems  +49 8669 31-3104 E-mail: service.ms-support@heidenhain.de NC support  +49 8669 31-3101 E-mail: service.nc-support@heidenhain.de NC programming ...

Table of Contents