Troubleshooting
Check Items When Trouble Occurs

An overload message (OVLD) is displayed (when nothing is connected to the

UNKNOWN terminal).
This is normal operation. An overload is often detected by the E4981B when
nothing is connected to the UNKNOWN terminal.

Beeping persists when turning ON the comparator function.

The front panel keys are unavailable.

o
If the keys have been locked through the handler interface, you cannot
unlock them by using the front panel keys. Set the /KEY_LOCK signal of the
handler interface to HIGH.

The measured value is abnormal.

Saving to USB memory fails

Some USB memory does not work with the E4981B. For more information on
USB memory, refer to

An error message or warning message is displayed.

For information on error messages and warning messages, refer to
D.
174
Check if the limit range is set up properly.
When the beep is set up to sound when the comparator function is ON
(initial setup), the beep sounds on every completion of measurement if the
limit range is set up improperly.
Check if the keys are locked.
When the keys are locked, LOCK is displayed in the status display area in
the lower-right corner of the screen.
Press the [Local/Lock] key to unlock the keys.
Check if the instrument is in the remote mode.
If the E4981B is in the remote mode, RMT is displayed in the status display
area in the lower-right corner of the screen.
Press the [Local/Lock] key to clear the remote mode.
The measurement of correction data may have failed.
Measure the correction data again. For how to measure the correction data,
refer to
"Obtaining Correction Data" on page
When the MULTI correction is ON, check that the channel and the definition
method of the LOAD standard value are selected correctly.
"USB interface port" on page
Keysight E4981B 120 Hz/1 kHz/1 MHz Capacitance Meter
98.
193.
Appendix

Advertisement

loading
Need help?

Need help?

Do you have a question about the E4981B and is the answer not in the manual?

This manual is also suitable for:

E4981b/001E4981b/002