12 Skin Temperature
12.1 Principle
The temperature probe measures the skin temperature by infrared temperature sensing.
12.2 The Skin Temperature Probe
Infrared temperature measurement is a non-contact principle. Ac-
cordingly, the probe should not be placed on the skin but kept at a
distance of approx. 1 – 3 cm.
To facilitate the measurement procedure an optional ventilating
spacer is available.
12.3 The Skin Temperature Screen
Up to eight temperature measurements can be made with continuous update of the average for
each measurement. Apply the probe to the skin and press Measure for each measurement.
It is possible to change between Fahrenheit and Celsius temperature scales.
12.4 Maintenance
The temperature probe is maintenance-free.
The temperature screen.
CORTEX TECHNOLOGY ApS