As .cmf files can be edited and files in general can become corrupted, it may be advantageous to
store files in a safe location and potentially create backups if needed.
10.6 Maintenance
For normal cleaning use a dry lint free cloth. For more thorough cleaning the cloth can be soaked
in a mild liquid detergent (soap). During cleaning of the probe, be careful not to introduce dirt or
liquids into the opening in the probe head as any dirt in the internal parts may adversely affect de-
vice performance. Solvents should not be used on the display window as it may become less
transparent and crack. When the device is not in use, it should be placed in its receptacle on the
®
DermaLab
Mini with the cable facing up to avoid any dust settling in the light guide.
10.6.1 Cleaning of the calibrator
For optimum performance, it is important to keep the working standards clean. Dust and other
loose particles can be blown off the surface using an inert dusting gas. Do not touch the surface
with your fingers or use your mouth to blow on the surface. If blowing off the surface is not
enough, a clean cotton wipe or lens tissue soaked in a solvent can be used. Usable solvents are
acetone, methylated spirit, and isopropyl alcohol (isopropanol). Read the safety data sheet care-
fully before using any solvents.
10.6.2 Cleaning of the measurement head
The measurement head can be cleaned and disinfected on the outer surface using an alcohol
swab or cloth soaked in a disinfectant if needed. Do not insert a cotton swab (or any other object)
into the measurement head.