9.8.3 Sliding door
Cleaning the sliding door
Fig. 98: Sliding door
9.9
Measures after maintenance is complete
2011-08-30
If the grooves for the sliding door (fig. 98/1) are noticeably soiled, they have to
be cleaned to ensure that the sliding door closes securely.
Personnel:
Trained person
Protective equipment:
1. Using a soft, dry cloth, clean the sliding door grooves (fig. 98/1).
2. Lightly oil the grooves.
The sliding door can seal shut properly and securely.
After completing maintenance work and before switching on the system,
perform the following activities:
1. Check all previously loosened screws to make sure they are secure.
2. Check that all previously removed protective devices and covers have been
properly reinstalled.
3. Ensure that all tools, materials and other equipment used has been cleared
away from the work area.
4. Clean the work area, removing any substance spillages (liquid or similar).
5. Ensure that all the system's safety devices are fully functional.
X-ray inspection system phoenix x|aminer
Protective gloves
Protective goggles
Protective clothing
Safety shoes
Maintenance
85