X-ray inspection system phoenix x|aminer
Maintenance
9.8.2 Pane of lead glass
Cleaning the pane of lead glass
Fig. 97: Pane of lead glass
Examining for damage
84
NOTE!
Material damage caused by use of incorrect detergent!
The high lead oxide content of the pane of lead glass makes
it highly susceptible to scratches, acids and lyes. Using the
wrong detergent may cause permanent damage to the
pane of lead glass and impair its protective function.
•
To clean it, use only water and a mild detergent.
•
Wipe the glass with a soft cloth only.
Personnel:
Trained person
Protective equipment:
1. Using a soft cloth and mild detergent, wipe the pane of lead glass (fig. 97/1)
2. Examine the pane of lead glass for damage.
3. If damage is detected, cease beam operation immediately and secure the
system against being switched back on. Notify the radiation protection adviser
and, if necessary, also the manufacturer of the X-ray inspection system.
Protective clothing
Safety shoes
2011-08-30