9.8
Maintaining and checking the radiation protection cabinet
9.8.1 Housing
Inspecting housing for damage
2011-08-30
If the radiation cabinet becomes noticeably soiled, you should perform the
following minor cleaning activities only:
Pane of lead glass: Section 9.8.2 "Pane of lead glass"
Sliding door: Section 9.8.3 "Sliding door"
Before the system is used, the radiation protection cabinet must be in-
spected for damage:
Housing: Section 9.8.1 "Housing"
Pane of lead glass: Section 9.8.2 "Pane of lead glass"
To guarantee safety for operating personnel, before the system is used, the
housing of the radiation protection cabinet must be inspected for damage.
Personnel:
Trained qualified personnel
Protective equipment:
1. Examine the radiation protection cabinet for external damage.
2. Shut down the system PC.
3. Switch off the X-ray inspection system and secure it against being switched
back on ( Section2.7 "Securing against the system being switched back on" ).
4. Open the left-hand service doors and then open the radiation pro-tection
door ( Section9.3 "Left-hand service doors" ).
5. Examine the interior of the X-ray protection cabinet for damage to the lead
shield.
6. If damage is detected, cease beam operation immediately and secure the
system against being switched back on. Notify the radiation protection adviser
and, if necessary, also the manufacturer of the X-ray inspection system
immediately.
X-ray inspection system phoenix x|aminer
Protective clothing
Safety shoes
Maintenance
83