3 Safety Information
Read the safety information below, before installation.
3.1 Warning
The SADPmini is not Intrinsically Safe. Therefore, it should not be used in hazardous
areas.
The maximum pressure to which the telescopic measuring head is exposed
0.3barg, 4.35psig, 30kpag or 0.3kg/cm
Exposing the measuring head to higher pressures may damage the
instrument and result in injury to the operator or other personnel in the area
3.2 Battery
3.2.1 Lithium-Ion
The SADPmini contains a Lithium-Ion battery which, to conform to current legislation
should always be partially discharged before shipping.
3.2.2 Initial Charging
• It is recommended that the SADPmini be put on charge overnight when first received.
Only use the Charger supplied with SADPmini
• To attain maximum battery life the SADPmini should always be re-charged at room
temperature.
• The battery life (hours of use from a fully charged cell) will reduce when used in cold
conditions. * 250 hours is quoted as the expected battery life when used at room
temperature.
Note: The 4-20mA output should not be operated while the instrument is charging.
must not be more than
Page 2 of 33