20 Instructions for Replacing Desiccant in the SADPmini
1. Fully extend the Desiccant head (1) of the SADPmini.
2. Unscrew the Retaining Collar (2), turning it counter clockwise (anti-clockwise),
which will unfasten the desiccant filled head from the main body of the SADPmini.
3. Gently pull out the Desiccant Head from the main body of SADPmini.
4. Remove the Desiccant Head Base "O" Ring (3) and keep in safe place.
5. Holding the Desiccant Head with the Desiccant Base Assembly (4) on top, fully
unscrew the Desiccant Head Base Assembly turning it counter clockwise (anti-
clockwise). The Desiccant Head base may require slight loosening by placing in a soft
jaw vice before this step is carried out. Do not fully unscrew with Head Base at the
bottom, as this will spill desiccant on floor.
USE TOOLS OR VICE WITH HARD
6. Dispose of used desiccant and re-fill with new up to a level of 3mm (5) from the
bottom of the screw thread.
7. Gently insert the Desiccant Head Base Assembly in the desiccant and tighten.
SOFT JAWS - DO NOT USE TOOLS OR VICE WITH HARD
8. Fit Desiccant Head Base "O" Ring (3)
9. Insert Desiccant Head in the main SADPmini body and tightly screw in the Retaining
Collar.
ONLY USE SOFT JAWS. DO NOT
JAWS.
Page 29 of 33
USE
JAWS.
Need help?
Do you have a question about the SADPmini and is the answer not in the manual?