Table 1-B. Accuracy to comparator method within Biosensor glucose ranges
(adults; n=130)
Biosensor
Number of
glucose
paired
1
range
Biosensor
(mg/dL)
YSI
<70
1,807
70–180
10,902
180–250
3,301
>250
4,320
1
Biosensor readings are within 40-400 mg/dL, inclusive.
Table 1-C. Accuracy to comparator method within comparator method glucose
ranges (adults; n=130)
YSI
Number of
glucose
paired
range
Biosensor
(mg/dL)
YSI
<70
2,008
70–180
10,463
180–250
2,988
>250
4,871
1
Biosensor readings are within 40-400 mg/dL, inclusive.
16 • Technical Information |
Percent
Percent
within 15
within 20
1
-
mg/dL YSI
mg/dL YSI
94.6
89.0
---
---
---
---
---
---
Percent
Percent
within 15
within 20
1
-
mg/dL YSI
mg/dL YSI
97.3
93.5
---
---
---
---
---
---
Stelo Glucose Biosensor User Guide
Percent
Percent
within 40
within
mg/dL YSI
15% YSI
99.5
---
---
84.8
84.3
---
91.6
---
Percent
Percent
within 40
within
mg/dL YSI
15% YSI
99.9
---
---
86.2
88.0
---
86.1
---
Percent
Mean
Percent
within
within
40% YSI
20% YSI
(mg/dL)
---
---
93.3
99.6
91.5
100.0
-11.3
96.8
100.0
Percent
Mean
Percent
within
within
40% YSI
20% YSI
(mg/dL)
---
---
93.8
99.7
94.9
99.9
92.5
99.9
-15.8
MARD
bias
(%)
-3.5
10.5
-1.1
8.5
8.3
-9.2
6.8
MARD
bias
(%)
2.3
10.5
-0.3
8.2
-6.5
7.4
8.0