Accuracy
The Stelo Glucose Biosensor System accuracy was assessed with paired sensor
readings to YSI blood glucose values. For YSI blood glucose values less than 70
mg/dL, the absolute difference in mg/dL between the two glucose results was
calculated. For values greater than or equal to 70 mg/dL, the absolute difference (%)
relative to the YSI value was calculated. In addition, the mean absolute relative
difference (MARD) shows the average amount the sensor readings differ from the YSI
glucose. The percentages of total readings within 20 mg/dL or 20% (20/20%) are
provided in Table 1-A. The tables are further categorized within Biosensor glucose
ranges (Table 1-B) and categorized within YSI glucose ranges (Table 1-C). When you
see a sensor reading on your mobile app, these tables show you how likely it is that
the reading matches your blood glucose level (measured by YSI in the study).
For example, the total number of data pairs considered in the analysis was 20,330. Of
these, 94.0% of the sensor readings fall within ± 20 mg/dL of the YSI blood glucose
values < 70 mg/dL and within ± 20% of YSI blood glucose values ≥ 70 mg/dL.
Table 1-A. Overall accuracy to comparator method (n=130)
Number of
Number of paired
Percent within
Percent within 20/20%
MARD
Population
1
subjects
Biosensor
-YSI
20/20% YSI
YSI on Day 1
(%)
Adults
130
20,330
94.0
88.6
8.3
1
Biosensor readings are within 40-400 mg/dL, inclusive.
90