Essential Performance - Dexcom stelo Manual

Advertisement

Model
Modulation
Data Rate
Data
Communication
Range
The maximum surface temperature of Applied part = 109.4°F.

Essential performance

The Stelo Glucose Biosensor System measures users' glucose sensor readings with
specified accuracy under the stated operating conditions. The Stelo Glucose Biosensor
System's essential performance also includes reporting the corresponding measured
glucose sensor readings on the display device.
Quality of service summary
Quality of service for the Stelo Glucose Biosensor System's wireless communication
using Bluetooth Low Energy is assured within the effective range of 20 feet,
unobstructed, between the Stelo Glucose Biosensor transmitter and paired display
device at regular 5-minute intervals. If connection is lost between the transmitter and
display device, upon re-connection any missed packets (up to 24 hours) will be
transmitted from the transmitter to the display device. The Stelo Glucose Biosensor
System is designed to only accept radio frequency (RF) communications from
recognized and paired display devices.
100
Stelo Glucose Biosensor
System
sensor/transmitter
Gaussian Frequency-Shift
Keying
1 Mbps
20 feet