Reference
Sequences (Adding and Deleting)
EXEC
unassigned
Sequences are automated routines that can perform tests at pro-
grammed times. The number of sequences that can be stored is lim-
ited only by the instrument's available non-volatile memory; adding
sequences reduces that memory. Executing sequences is described in
the Making Measurements section.
Adding Sequences
24 hour test. The 24 hour test sequence is installed in the RFM 150
at the factory and can not be deleted. Additional sequences can be
imported using either another RFM 150 or the CSS150 software
package.
Using another RFM 150. Connect the two instruments, using the
instructions on page C—2. On the source instrument, press SEQ to
display the list of sequences (Figure 3-11). Turn the knob to select
the sequence to be sent, then select CLONE SEQ. The sequence is
sent to the destination RFM 150 and placed at the end of the list of
sequences. To display this list on the destination monitor, press SEQ.
Use the knob to scroll if needed.
Using the CSS150 software package. Predefined sequences can
easily be selected and imported to the RFM 150, and other sequences
can be quickly set up by turning fields on or off with the mouse.
Refer to the CSS 150 User Manual for instructions.
Deleting Sequences
3-22
Delete all. To delete all sequences stored in the instrument, press
SEQ, then select DEL ALL. A confirmation screen will appear;
select YES to delete all sequences. Selecting NO or ESC will cancel
the request.