Setting Value Obtaining and Changing Command
Measurement Setting 1 (Setting per TASK)
Unit No.
Data No.
(Note 1)
28h
00h
01h
02h
03h
04h
05h
06h
07h
ZS Series
3-20
CompoWay/F Communication Command Reference
Parameter
Mode
Input selection
Expression
Calculation parameter X
Calculation parameter Y
Calculation parameter K
Calculation parameter m
Calculation parameter n
Setting range/output
range
0: OFF
1: INDIV
2: OPERATION
0: Input A
1: Input B
2: Input C
3: Input D
4: Input E
5: Input F
6: Input G
7: Input H
8: Input I
0: THICKNESS (K-(X+Y))
1: STEP (X-Y)
2: K+mX+nY
3: AVE
4: MAX-MIN
0: Input A
1: Input B
2: Input C
3: Input D
4: Input E
5: Input F
6: Input G
7: Input H
8: Input I
9: TASK 1
10: TASK 2
11: TASK 3
12: TASK 4
-999999999 to
999999999 (nm)
-100 to 100 (X0.1)
-100 to 100 (X0.1)
Remarks
Parameters when
[Mode]=[INDIV]
Parameters when
[Mode]=[Operation]
Parameters when
[Expression]=
[Thickness], [STEP],
and [K+mX+nY]
Parameters when
[Expression]=
[K+mX+nY]