6.14 UNIT-2
Value of a scale verification interval for the second interval. In this parameter you can set the value of the
scale division for second range for dual range scales or scale interval for single range scales.
The unit should be set according to point 4.3 - in grams, Kg. Pay attention to the decimal point.
Eligible values are 1,2,5,10 and the outputs of these numbers by 10,100, 1000, etc.
6.15 CNT-2
Number of verification scales intervals for second interval. The multiplication of "CNT-2" and "UNIT-2"
equals to the max scale capacity.
For legal for trade scales"CNT-2"≤3000.
6.16 TRADE
If parameter is set to =1, no weight values will be sent through the interfaces is weight is below minimum.
6.17 DPOINT
Number of digits after the decimal point.
In mode Kg possible values are: 0, 1, 2, 3
In model Grams possible values are 0, 1, 2
6.18 O-LOAD
Number of verification scales intervals above the scale maximum, eligible for display.
For legal for trade versions parameter should be set to ≤ 9.
6.19 U-LOAD
Minimum number of verification scale intervals (Min) under which some functions are not allowed.
Functions are: Sending weight through interface, Auto tare, Print.
For legal for trade scales values should be 20
6.20 U1-
A
R
N
Max capacity for Range 1 in kg
6.21 U2-
A
R
N
Max capacity for Range 2 in kg
6.22 GRAV-1
Value of gravity coefficient for the place of calibration.
6.23 GRAV-2
Value of gravity coefficient for the place of use.
Note: If some of the parameters "GRAV-1" or "GRAV-2" is set to zero, no gravity compensation will be
active.