6.8 UCAL-2
Value of the second calibration point in unit according to set up in section 4.3. Grams, Kg.
!!!Pay attention to the decimal point in mode grams/kg!!!
This calibration point in mandatory! Enter the calibration weight and press
parameter.
6.9 DELTA2/CAL-2
Shows the calibration weight for second calibration point.
The number is shows in ADC values minus the ADC value of Zero point. Refer to section 4.5.
Load the scale with the corresponding load set in parameter "WCAL-2". Wait for the load to get stable (3-5
sec) press and hold key
Display will show "GET-1/CALIB", after which the display shows the new ADC value stored for this point.
Press
go to the next parameter.
6.10 STB-1
Number of successive reports tracked when determining whether the balance is in a stable state.
Normal value is 5 reports.
6.11 STB-2
Allowable difference between two successive reports when determining whether the scale is in a stable
state. The unit of measurement is parts of the division (see next point) of the scale.
The normal value is 0.35 parts of the division.
The parameters "STB-1" and "STB-2" influence the operation of the balance when exchanging data with
other devices when printing, with automatic functions (automatic tare and auto-lock). Very small values of
"STB-1" can compromise the mentioned functions, very large values - delay these functions.
Conversely, large values of "STB-2" may compromise the mentioned functions, very small values - delay
these functions
6.12 UNIT-1
Value of a scale verification interval in the first interval. In this parameter you can set the value of the
scale division for first range.
In the case of a single range scale, this parameter must be set to 0.
The unit should be set according to point 4.3 - in grams, Kg. Pay attention to the decimal point.
Eligible values are 1,2,5,10 and the outputs of these numbers by 10,100, 1000, etc.
6.13 CNT-1
Number of verification scales intervals for first interval. The multiplication of "CNT-1" and "UNIT-1" equals
to the capacity of the first weighing range.
For legal for trade scales"CNT-1"≤3000.
.
to go to next