5. SERVICE MENU
To enter service settings from menu
Hold
until display shows SETUP
Confirm with
and choose menu SYSTEM as shows blow
Scrow to main setup menus »
INTERF »
» SYSTEM»
SYSTEM menu has the following sub menus:
PARAMS – Scale general parameters (jumper protected)
DEAFAUL LOAD – Default factory stetting Reset
SYST-2 – LCD display brightness adjustment
CALIB POINT – Detailed calibration menu (jumper protected)
CALIB CPARAM – Advance Calibration parameter for Certification purposes (jumper protected)
SYS SPEED – System speed parameters
6. DETAILED CALIBRATION ( CALIB POINT )
Calibration menu navigation:
The numeric filed can be changed by the numeric keys or arrows UP an RIGHT.
Key
used to save the changes and go to the next parameter.
Key
used to take scale Zero
Key
use to take calibration
Key
used to to go to next parameter and exit/save calibration
6.1 RANGE_
When changing RANGE parameter and confirming with ENTER – indicator checks if there is such table in
the memory. If such table is existing it loads all parameters. If not tables is found all remaining parameters
remain unchanged.
Note: In current version only table 0615 is available.
» ARTICL »
» FUNCS »
» PRINT »
»