Elicom Electronic S300F Series Service Manual

Hide thumbs Also See for S300F Series:

Advertisement

Quick Links

SERVICE MANUAL
WEIGHING SCALE
SERIES S300F, S300L
July 2020
Version 1.0

Advertisement

loading

Summary of Contents for Elicom Electronic S300F Series

  • Page 1 SERVICE MANUAL WEIGHING SCALE SERIES S300F, S300L July 2020 Version 1.0...
  • Page 2 1. INTRODUCTION The scale handles the weight according to data stored in a memory table during the calibration. When setting up the balance, you should specify a number of parameters described below. When re-calibrating, it is enough to take a point at zero and one or two points with a load. At initial setup, it is convenient to load default data, which can then be changed according to the type and scope of the scale.
  • Page 3: Enter Calibration Mode

    3. ENTER CALIBRATION MODE Before you enter calibration, the jumper switch should be activated to allow parameter change. 1. Remove the calibration cover to access the calibration switch. 2. Press the calibration switch (jumper) while scale is operating. Scale display will show a brief message “Jumper On” 3.
  • Page 4 4.8 Press to go to the calibration of the first calibration point CAL-1 / Delta2. 009472 CAL1/DELTA1 4.9 Place the calibration test weight in the middle of the scale pan. In our Example 2kg. If you use single point calibration skip this step by pressing 4.10 To take the calibration press Message GET-1 CALIB will appear to confirm the calibration...
  • Page 5 The scale has built in function for gravity compensation in cases where the place of calibration differs from place of use. Parameter GrAv-1 is the G-value for the place of calibration, where parameter GrAv-2 is the G- value of the place of use. In cases where the G-value of calibration is same as G-value of use. Parameter GrAv-2 should be set to 0.
  • Page 6: Service Menu

    5. SERVICE MENU To enter service settings from menu Hold until display shows SETUP Confirm with and choose menu SYSTEM as shows blow Scrow to main setup menus » » ARTICL » » FUNCS » » PRINT » » INTERF » »...
  • Page 7 6.2 SERNUM Scale/module serial number. It can be changed only it its 0000000. 6.3 G-KG Parameter for main weight unit. 0 – Grams 1 – Kg 6.4 ADC___ Shows current ADC reading from ADC module. Used to evaluate the scale condition and loading/unloading reaction. No a parameter for change/set.
  • Page 8 6.8 UCAL-2 Value of the second calibration point in unit according to set up in section 4.3. Grams, Kg. !!!Pay attention to the decimal point in mode grams/kg!!! This calibration point in mandatory! Enter the calibration weight and press to go to next parameter.
  • Page 9 6.14 UNIT-2 Value of a scale verification interval for the second interval. In this parameter you can set the value of the scale division for second range for dual range scales or scale interval for single range scales. The unit should be set according to point 4.3 - in grams, Kg. Pay attention to the decimal point. Eligible values are 1,2,5,10 and the outputs of these numbers by 10,100, 1000, etc.
  • Page 10 Exit Calibration mode To exit calibration mode hold Hold until display shows WEIGHT SAVE. To save settings confirm with Instrument will restart. Test to see if calibration changes are correct and seal the instrument. To exit without saving the changes press display shows WEIGHT SAVE confirm with 7.
  • Page 11 7.7 GRAV-2 Value of gravity coefficient for the place of use. Note: If some of the parameters ”GRAV-1” or ”GRAV-2” is set to zero, no gravity compensation will be active. 7.8 USEC-Z If parameter set =1, when turning ON the instrument, the indicator will not initial zeroing and calibration Zero will be used.
  • Page 12 Maximum weight to be displayed below ZERO in number of verification intervals (see section 4.36) For Legal for trade scales ≤19 7.16 HWGAIN ADC parameter. SET by Default to 0 7.17 T-COR Temperature compensation. 1 – Disabled 0 – Enabled Table 1.
  • Page 13 STB-1- JD3.0 Count of sequential readings, to set stability flag STB-2- 00.40 JD4.2 Allowed difference between sequential readings to set stability flag. TRAC-0 1.50 JD3.2 In ‘e1’ (‘e’) units – Zero autotrack range. TRADE_ JD1.0 Flag 1/0 – scale is in trade mode. DPOINT JD1.0 Place of decimal point.
  • Page 14 T-cor_ Jb1.0 Enabled(1)Disabled(0) – temperature compensation G-KG Jb1.0 Main unit 0 = g 1 = kg  R -Read only  J - write by Jumper  W - write w/o jumper  D - number  b -bitflag...

This manual is also suitable for: