1. INTRODUCTION
The scale handles the weight according to data stored in a memory table during the calibration.
When setting up the balance, you should specify a number of parameters described below.
When re-calibrating, it is enough to take a point at zero and one or two points with a load.
At initial setup, it is convenient to load default data, which can then be changed according to the type and
scope of the scale.
The balance uses 2 or 3 calibration points - one for unloaded position (zero point) and one or two for load
(load points). It is mandatory to use zero and second load point, first load point is not mandatory, but
must be set to 0 if not used.
2. View from below
5
1
Fig. 2.1 – View from below S300
Fig. 2.2 – View from below S300-M
4
Table 2.1 – View from below S300
3
Table 2.26 – View from below S300-M
1.
Power supply
connection
2.
RS232 interface
connection
3.
Calibration seal
4.
Battery
compartment
5.
Adjustable feet
1.
Power supply
connection
2.
RS232 interface
connection
3.
Calibration seal
4.
Adjustable feet