Elicom Electronic S300F Series Service Manual page 7

Hide thumbs Also See for S300F Series:

Advertisement

6.2 SERNUM
Scale/module serial number. It can be changed only it its 0000000.
6.3 G-KG
Parameter for main weight unit.
0 – Grams
1 – Kg
6.4 ADC___
Shows current ADC reading from ADC module.
Used to evaluate the scale condition and loading/unloading reaction.
No a parameter for change/set.
6.5 ADC-0-
Shows ADC value of calibration point ZERO.
To take calibration zero, ensure that scale is free of weight and hold
Display will shows "GET-0-/CALIB" for about 2 sec., after that it will remember the new ADC value for
ZERO.
6.6 UCAL-1
Value of first calibration point in unit according to set up in section 4.3. Grams, Kg.
Pay attention to the decimal point in mode grams/kg.
If calibration will be made at single calibration point, calibration point UCAL-1 should be set to zero
(000000).
6.7 DELTA1/CAL-1
Shows the calibration weight for first calibration point.
The number is shows in ADC values minus the ADC value of Zero point. Refer to section 4.5.
If first calibration point will not be used ("WCAL-1"= 000000), press
If first calibration point will be used ("WCAL-1" > 000000), load the scale with the corresponding load set
in parameter "WCAL-1". Wait for the load to get stable (3-5 sec) press and hold key
Display will show "GET-1/CALIB", after which the display shows the the new ADC value stored for this
point.
Press
go to the next parameter.
key.
to go to next parameter.
.