4.8 Press
to go to the calibration of the first calibration point CAL-1 / Delta2.
4.9 Place the calibration test weight in the middle of the scale pan. In our Example 2kg.
If you use single point calibration skip this step by pressing
4.10
To take the calibration press
Message GET-1 CALIB will appear to confirm the calibration
this is the first weight calibration weight point. If you wish to make a single point calibration set
uCal-1 to 00000.0
4.11
Press
Enter second calibration point.
In the example we will calibrate second point with 5kg. Enter 00005.0 using the numeric keyboard.
4.12
Press
4.13
Place the calibration test weight in the middle of the scale pan. In our Example 5kg.
4.14
To take the calibration press
Message GET-2 CALIB will appear to confirm the calibration
4.15
Press
009472
CAL1/DELTA1
GET-1
CALIB
to go the second calibration parameter uCal-2.
00005.0
uCAL-2
to go to the calibration of the second calibration point CAL-2 / Delta2.
023680
CAL2/DELTA2
GET-2
CALIB
to go to the Gravity setting parameters.
.
.