The scale has built in function for gravity compensation in cases where the place of calibration
differs from place of use.
Parameter GrAv-1 is the G-value for the place of calibration, where parameter GrAv-2 is the G-
value of the place of use. In cases where the G-value of calibration is same as G-value of use.
Parameter GrAv-2 should be set to 0.
4.16
GrAv-1 parameter is set by default to g=9.8053 as factory G-value
You can change the value by the numeric keyboard if needed. And confirm by
098053
GrAv-1
4.17
GrAv-2 parameter is g-value for the place of use.
If gravity compensation is not necessary parameter should be set to 0 or you should enter the g-value
for the place of use and confirm it by
000000
GrAv-2
4.18
After Calibtaiton is done, hold
WEIGHT
SAVE
Press
to save the calibration
4.19
Scale will restart automatically and new calibration will be saved.
until message WEIGHT SAVE appear.