Enter Calibration Mode; Quick Calibration - Elicom Electronic S300F Series Service Manual

Hide thumbs Also See for S300F Series:

Advertisement

3. ENTER CALIBRATION MODE

Before you enter calibration, the jumper switch should be activated to allow parameter change.
1. Remove the calibration cover to access the calibration switch.
2. Press the calibration switch (jumper) while scale is operating.
Scale display will show a brief message "Jumper On"
3. When Jumper ON, the scale will show high resolution x10 weighing with icon "x", and G-value
compensation will not be active
4. Scale is ready for calibration

4.1 QUICK CALIBRATION

Quick calibration mode allows only scale re-calibration and G-value adjustment.
4.1 To access calibration mode hold
4.2 Once in calibration mode, display will show ADC reading
You can load the scale to the is the load cell reading is active and stable
This parameter is accessible without Jumper and is use to check load cell and ADC performance
4.3 In calibration mode use the
4.4 Press
to go to parameter ADC-0-, scale calibration Zero.
4.5 Take calibration Zero by holding
4.6 Press
to go to the next calibration parameter uCal-1, this is the first weight calibration weight
point. If you wish to make a single point calibration set uCal-1 to 00000.0
4.7 Enter first calibration point.
In the example we will calibrate first point with 2kg. Enter 00002.0 using the numeric keyboard.
086585
ADC
or
086585
ADC-0-
Get -0-
Calib
00002.0
uCAL-1
key for 3 sec.
to go through calibration parameters.