Exit Calibration mode
To exit calibration mode hold
Hold
until display shows WEIGHT SAVE.
To save settings confirm with
Instrument will restart. Test to see if calibration changes are correct and seal the instrument.
To exit without saving the changes press
7. CALIB RAPAMETERS
7.1 TRAC-0
Scope - in parts of the division - to activate automatic zero-setting device.
Normal value is 1.50.
7.2 TOL-0
Permissible deviation / tolerance at initial reset - in percent of scale range. When the instrument is
switched on, the initial reset monitors whether the measured load (relative to zero of the calibration) is
less than or greater than this tolerance. If larger, it issues an error-0- message and remains in the initial
reset state.
Allowed value for legal for trade scales ≤ 10%.
Secondary zeroing by key or automatic zero-setting device is limited to 4% of the range fromt he zero
setting point.
7.3 NEAR-0
In parts of the verification scale interval - Determines the action of the proximity indicator to zero.
Value for legal for trade scales is 0.30
7.4 FAST
An assist parameter, indicating:
-
slow operation of the ADC (0,1,2,3)
-
fast operation of the ADC (10, 11, 12, 13)
-
fast operation with hitting peaks (20, 21, 22, 23).
7.5 TCAL_
Measured temperature at calibration. The value is not an accurate temperature and it is used for
temperature deviations and weight calculation.
7.6 GRAV-1
Value of gravity coefficient for the place of calibration.
.