19 Troubleshooting
19.2 Error Messages
Overheat
Code
1036
1037-1044
An error may have occurred in the temperature control circuit.
Turn off the system, and contact your Shimadzu representative.
Relay/heater control errors
Code
1045
1046
1047
1048
An error may have occurred in the control circuit. Turn off the system, and contact your
Shimadzu representative.
Code
1049-1052
This message appears when the Detector temperature setting value is smaller than that of
the Oven.
If this message appears, change the setting value of the Detector temperature.
If set the setting of Protection against contamination to "No", need not to change the setting
value of the Detector temperature. (Refer to 16.6.4.3)
Pressure/flow rate control errors
Code
2005-2007
2014, 2017
2020, 2023
2015, 2018
2021, 2024
2016, 2018
2022, 2025
2026-2043
The prressure cannot reach the set value. Check whether gas is supplied and whether gas
is leaking from connections.
NOTE
Gas leaks may be occurring in several locations.
(Example) If the purge flow rate is too low,the message "Purge leaks" may appear. If the carrier gas is
GC-2010 Plus
324
Message
Overheat is sensed
xxx temp controller broken
Message
COL relay error
INJ1-DET2 relay error
AUX3-AUX5 relay error
Heater controller error
Message
DET#n set temp is low
Message
CARx purge leaks
DetAPC n makeup gas leaks
DetAPC n hydrogen leaks
DetAPC n air leaks
APC y leaks
also leaking, the ESC or TFC error message may not be displayed first.
CS
Note
Overheat detected by
Yes
thermocouple
xxx=COL, INJ1, DET1,
Yes
INJ2, DET2, AUX3,
AUX4, AUX5
CS
Note
No
No
No
No
CS
Note
No
n=1-4
CS
Note
Yes
x=1-3
Yes
Yes
n=1-4
Yes
Yes
y=1-18