12 Injection Port
12.6 Direct Injection System
12.6.1

Setting the temperature

12.6.1.1
Screen description
Press the [INJ] key to display the injection port main screen shown in Fig. 12.6.2 and to
set the injection port temperature.
12.6.1.2
Parameter list
TEMP
Range: 0.0−450.0 °C, Default: 25.0 °C
Set the injection port temperature.
Keep the injection port temperature as low as is practical to prevent of a decrease in the
service life of the temperature sensor. To instantly vaporize the injected sample, set the
injection port temperature which is suitable for the sample. Because the sample quantity
is usually minute, the sample is vaporized at temperatures lower than the boiling point
of sample compounds. To set the maximum temperature limit, refer to
the maximum temperature limits"
12.6.1.3
PF menu
PF menu
Print
Next Inj
GC-2010 Plus
142
Fig. 12.6.2   [INJ] key main screen
Prints the temperature program to a Chromatopac.
Toggles to the next setup screen when two or more injection ports are
installed in this system.
The [INJ] key can also be used to switch screens.
in "16.6 GC Configuration".
Description
"16.6.4 Setting
Reference
section
――
――