Technical Data - Xylem wtw Cl 800 DIN Operating Manual

Table of Contents

Advertisement

Available languages

Available languages

Cl 800
Measuring ranges
Reproducibility
pH range
Temperature range
Resistance
Length
Diameter
Cable length
Plug
ba15355d04
12/2021
Cause
– Inner junctions encrusted
– Electrode defective

Technical data

-
2 ... 35,000 mg/l Cl
± 2 %
2 ... 12
(see section I
0 ... 80 °C
(shortly up to 100 °C)
< 1 MΩ
153 mm (including 33 mm connection head)
Shaft:
Connection head:
1 m
DIN plug
Remedy
– Leave the bridge electrolyte to react on
the inner junctions for some hours until
the crusts have dissolved.
– Exchange the electrode
-5
-
(10
... 1 mol/l Cl
)
NTERFERENCES
12 mm
16 mm
Technical data
)
23