Clinician Guide – Page 36
A-36
ENGLISH
13.2 Error Messages on SleepStyle Screen
If a fault is detected with the device or its accessories, a screen will appear with a notification and an error
code. Identify the error code in the ranges specified below and follow the appropriate corrective action. If
the error continues or reoccurs, record the number displayed and contact your Fisher & Paykel Healthcare
representative for further instructions.
Error codes
Solution
between
100–199
The device may not be able to provide effective therapy. The device may have shut down
or may not be able to provide the prescribed pressure.
Solution: Turn the power off and on at the power supply to restart the device.
400–499
Humidity may have been disabled. The device is still safe to use without humidity. The
patient will still be treated at the prescribed pressure.
Solution: Turn the power off and on at the power supply to restart the device.
510 or 512
There may be a problem with the ThermoSmart breathing tube. The device is still safe to
use. The patient will still be treated and get humidity, but it may not be optimal.
Solution: Try reconnecting your ThermoSmart breathing tube. When connected
correctly, the ThermoSmart icon
the power off and on at the power supply to restart the device. If the error continues or
reoccurs, replace the ThermoSmart breathing tube. See section 12 - Replacement Parts.
500–599
The ThermoSmart breathing tube may have been disabled. The device is still safe to use.
(excluding 510
The patient will still be treated and get humidity, but it may not be optimal.
or 512)
Solution: Turn the power off and on at the power supply to restart the device. If the error
continues or reoccurs, See section 12 - Replacement Parts.
will appear on the home screen. Alternatively, turn
F&P SleepStyle Product Technical Guide | 78