GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The F&P SleepStyle Series is intended for use in the electromagnetic environment specified below. The customer or the user of F&P SleepStyle
Series should ensure that it is used in such an environment.
Immunity test
IEC60601 test level
Conducted RF
6 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
a
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 6 V/m.
b
Compliance level
6 Vrms
10 V/m
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the device, including
cables, than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
d = 0.58 √P
d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
should be less than
a
the compliance level in each
frequency range.
Interference
b
may occur in the vicinity of
equipment marked with the
following symbol:
F&P SleepStyle Product Technical Guide | 12

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SleepStyle Auto and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Sleepstyle cpap

Table of Contents

Save PDF