Clinician Guide – Page 10
A-10
ENGLISH
4. THERAPY OPTIONS
4.1 Therapy Mode
In this menu, you can select the pressure mode.
This can be changed from CPAP (fixed pressure),
to Auto (auto-adjusting pressure) in device models
with this feature (F&P SleepStyle Auto only).
1.
Press Down
or Up
"Therapy mode" menu.
2. Press OK
to select "Therapy mode".
3. Press Down
or Up
selection.
4. Press OK
to confirm your selection.
Note: Pressing OK
will automatically take you
to set the therapy pressure. Pressing Menu
take you back to the main Clinician Menu.
4.2 Setting the Therapy
Pressure
4.2.1 CPAP mode
This setting allows the prescribed fixed pressure
to be set.
The device has a default fixed pressure setting of
10 cmH
O, which can be adjusted in increments of
2
0.5 cmH
O within the range of 4 to 20 cmH
2
1.
Press Down
or Up
prescribed pressure.
2. Press OK
to confirm and exit to the main
Clinician Menu.
4.2.2 Auto mode (Auto models
only)
Minimum pressure: Adjust the minimum starting
pressure in 0.5 cmH
to the maximum pressure setting. The default
setting is 4 cmH
to scroll to the
1.
Press Down
minimum pressure.
2. Press OK
to change the
maximum pressure.
Maximum pressure: Adjust the maximum pressure
in 0.5 cmH
pressure setting to 20 cmH
will
is 18 cmH
1.
Press Down
maximum pressure.
2. Press OK
O.
2
Clinician Menu.
Note: Minimum and maximum pressure can be
accessed independently from the main Clinician
Menu, under Therapy.
to change the
O increments from 4 cmH
2
O.
2
or Up
to change the
to confirm and move on to
O increments from the minimum
2
O. The default setting
2
O.
2
or Up
to change the
to confirm and exit to the main
O
2
F&P SleepStyle Product Technical Guide | 52