Clinician Guide – Page 9
Sub-menu
Device Options

Service Options

Settings
Expiratory relief
Activate Expiratory relief:
Activates or Deactivates the expiratory relief feature.
If No
is selected, expiratory relief will be Deactivated, you will
be taken back to the main Clinician Menu screen, and the "Expiratory
Relief" Patient Menu item will be removed from the Patient Menu.
Expiratory relief (level):
When expiratory relief is Activated, the next automatic menu screen will
allow you to adjust the expiratory relief level from 0 – 3 pressure units
(cmH
O).
2
Allow patient to adjust:
If Yes
is selected, expiratory relief will be visible in the Patient
Menu and the patient will be able to adjust the expiratory relief level
from Off (0 shaded circles) to On, and up to 3 (3 shaded circles).
If No
is selected, expiratory relief will disappear from the Patient
Menu so that the patient cannot choose to adjust the expiratory relief
level. Expiratory relief will still run in the background as set up by the
clinician.
Humidity
Adjust the level of humidity; the default setting is 5.
Display pressure
Option to show or hide pressure whilst therapy is on. Also set the
pressure units of the device.
Display AHI
Option to show or hide AHI when a patient views their therapy data.
Time
Set the time of the device's internal clock to ensure therapy data is
reported accurately (24 hour format).
Language
Set the language that you would like the device to display.

Information

View software version and total runtime hours of the device.

Fault log

A list of fault codes and the dates on which they occurred since last

factory reset.

Clear patient data

Clears all therapy data from the device and the InfoUSB linked to the
device.
Factory reset
A factory reset removes all data and prescription settings from the
device and InfoUSB. It will restart the device with the default factory
settings. To avoid data corruption, please do not turn off the device
while the factory reset is in progress. After a factory reset, the device
will restart as if it is being started for the first time. Refer to section 2.3 –
First Start-up.
ENGLISH
A-9
F&P SleepStyle Product Technical Guide | 51