Replacing The Air Filter Between Patients; Deleting Therapy Data Between Patients - F&P SleepStyle Auto Technical Manual

Hide thumbs Also See for SleepStyle Auto:
Table of Contents

Advertisement

Clinician Guide – Page 29
10.4 Replacing the Air Filter
between Patients
The air filter is located at the rear of the device
and ensures the air intake into the device is
filtered to remove particles and dust. The air filter
is not washable and must be replaced between
patients. To replace the air filter, please follow the
instructions below.
Warnings
To avoid choking, or inhalation of a foreign
object:
Do not use the device without the
recommended air filter fitted. This will reduce
dust or particles entering the device and
breathing tube.
1.
To remove, pinch the air filter with your fingers
and pull it out of the device.
2. Hold onto the short side of the new air filter.
Push into the device so there are no gaps.
10.5 Deleting Therapy Data
between Patients
For accurate compliance reports, data must be
cleared from the device memory, as well as the
InfoUSB, before use by another patient. Follow the
instructions in section 7.3 – Clear Patient Data to
delete therapy data from the device's memory.
ENGLISH
A-29
F&P SleepStyle Product Technical Guide | 71

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SleepStyle Auto and is the answer not in the manual?

Questions and answers

emma
February 14, 2025

is it possible to increase air pressure from machine into mask as it doesnt fully stop me snoring?

1 comments:
Mr. Anderson
February 14, 2025

The F&P SleepStyle Auto has an auto-adjusting pressure feature, which means it can automatically adjust air pressure based on the user's needs. However, the manual does not explicitly state that the pressure can be manually increased to reduce snoring.

This answer is automatically generated

Subscribe to Our Youtube Channel

This manual is also suitable for:

Sleepstyle cpap

Table of Contents

Save PDF