To Check The Line Filter Board Ac Output - Keysight Technologies InfiniiVision 3000G X Series Service Manual

Table of Contents

Advertisement

Test Point
Expected Value
(near J3701)
VP5V
+5 V ±3%
VP12V
+12 V ±3%
VP1V8
+1.8 V ±3%
VP1V4
+1.4 V ±3%
VP1V
+1 V ±3%
VP1V2
+1.2 V ±3%
VP2V5
+2.5 V ±3%
VP3V3
+3.3 V ±3%

To check the line filter board AC output

When the 13 V bulk power is not being properly supplied, this procedure tests the
AC input to the power supply to determine whether there is a problem with the line
filter board.
1 Follow the instructions in
2 Connect the power cord, and turn on the oscilloscope.
3 Verify that AC power is present at J1 on the power supply using a DVM and
3000G X-Series Oscilloscopes Service Guide
Assemblies
If Good
Supplied
Keyboard,
Go to
System Board
supplies"
System Board
If good when the interboard supply
cable is disconnected from the
System Board
system board, but bad when the
interboard supply cable is
System Board
connected, this indicates a problem
System Board
with either the cable or the
assemblies supplied,
System Board
1 First, test the interboard supply
System Board
2 If the interboard supply cable is
LCD, System
Go to
Board
supplies"
troubleshooting"
on page 70.
probes as shown in the following picture (remember this is an AC voltage
measurement!).
"To check the keyboard
on page 83.
cable for opens or shorts using
the DMM.
good, replace the system board.
"To check the display
on page 82.
"To prepare for internal assembly
Troubleshooting
If Bad
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
Replace the line filter board.
4
79