To construct the test connector (for use with MSO models only)

Keysight 3000G X-Series oscilloscopes that have digital channels enabled require
the test connector described below. Follow the steps to build the test connector.
Table 3
Description
BNC (f) Connector
Berg Strip, 8-by-2
Jumper wire
1 Obtain a BNC connector and an 8-by-2 section of Berg strip. A longer strip can
2 On one side of the Berg strip, solder a jumper wire to all of the pins (shown in
3 On the other side of the Berg strip, solder another jumper wire to all of the pins.
4 Solder the center of the BNC connector to a center pin on one of the rows on
5 Solder the ground tab of the BNC connector to a center pin on the other row on
3000G X-Series Oscilloscopes Service Guide
Materials required to construct the test connectors
be cut to length using wire cutters.
Figure 1
on page 24).
the Berg strip.
the Berg strip.
Recommended Part
Keysight 1250-1032 or
Pomona 4578
3M .100" x .100" Pin Strip
Header or similar
Testing Performance
Qty
1
1 strip, cut to length (8x2)
2
23