User Interfaces; Outputs - Siemens SITRANS L Compact Operating Instructions

Controllers with ultrasonic transducer inputs
Table of Contents

Advertisement

Technical specifications

8.3 User Interfaces

8.3
User Interfaces
Configuration
Display (local)
Connectors
8.4

Outputs

mA analog active output
Relays
Control relay
Alarm relay
Optional analog channel on the HART communi-
cation card
All relays are certified only for use with equipment that fails in a state at or under the rated
1)
maximums of the relays.
Orderable with 1, 3, or 6 relays
2)
104
Local buttons
Advanced graphical liquid crystal display
USB Service port
USB version: V2.0
USB socket: Mini-B
Accessory port
Single-point version includes one mA output
Dual-point version includes two mA outputs
1)
Six, maximum
2)
USB port used for
SIMATIC PDM/SITRANS DTM, FW up-
date, Data log extract
0 to 20 mA
4 to 20 mA
Accuracy
± 20 uA, over 3.5 to 22.6 mA
± 40 uA, below 3.5 mA
Resolution 3 uA, over 0.1 to 22.6 mA maxi-
mum
750 ohm maximum
Isolated (500 V DC)
4 control
2 alarm control
All relays rated 5A at 250 V AC, non-inductive
4 Form A, NO relay (numbers 1, 2, 4, 5)
2 Form C, NO, or NC relay (numbers 3, 6)
0 to 20 mA without HART
4 to 20 mA with HART:
Passive - 14 to 30 V (at the terminals),
500 ohm maximum
Active - 350 ohm maximum
SITRANS LT500 with ultrasonic transducer inputs
Compact Operating Instructions, 02/2024, A5E52021877-AA