Installing/mounting
4.2 Installing/mounting the device
4.2.5

Remote panel mounting

4.2.6

Mounting the remote lid

Note
Remote mounted lid can be mounted up to 5 m from the device using two optional cable
extensions (each 2.5 m in length). For instructions on how to connect an extension cable, see
Remote mounted lid with extension cable (Page 24).
1. Using the template provided, cut out the necessary hole for the remote display lid. Place the
gasket inside the lid, aligning the mounting holes. Align the back of the remote display lid
with the panel hole cutout. Mark and drill four holes in the mounting surface for the four
screws (provided).
2. Fasten with a screwdriver and wrench.
22
Four panel mount fasteners provided
Remote display lid
Gasket
LT500 with blank lid
Extension cable
Connection between two extension cables
SITRANS LT500 with ultrasonic transducer inputs