Activation - Crowcon Clip O2 User Manual

Portable single gas detector
Table of Contents

Advertisement

Clip SGD
CAUTION
Before use, check the activate by date on the box. If the activation date is past, do not
activate the device.
Shelf Life within 6 months from manufacture.
Only activate in a safe environment. Press the button for 3 secs; gas type and firmware
version (e.g. v2.2) will be displayed. Following the 10 secs countdown, the device will be
fully activated, the
Measurement mode.
After turning on the device, the remaining life is displayed; 23 months originally. After the
remaining life reaches 1 month, 30 days remaining is displayed. When 1 day is left, 24
hours(T) remaining is displayed.
D
CAUTION
Appropriate calibration of the device is required prior to operation. Always ensure that
the device responds correctly to the relevant gas. Verify that foreign materials that could
interfere with the detection of gas are not blocking the sensor.
icon will flash on the display screen and the device will be in Gas
V
1
ALARM
2
M
D
T
In the event that activation of the device fails, the
X
icon will appear on the display and Gas Measurement
mode will not be entered. In this case, perform calibration
or contact authorized reseller or Crowcon Detection
%
Instruments customer support at +44 (0) 1235 557711
for repair/return information.
V
X
MAX
M N
I
%
LEL
T
PPM
%
VOL

2. Activation

V
%
X
7

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Clip O2 and is the answer not in the manual?

This manual is also suitable for:

Clip sgdClo195

Table of Contents